Command reference
R&S
®
ZND
776
User Manual 1173.9557.02 ─ 61
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:EMBedding<LogPt>:
PARameters:L<Cmp>
<CircuitModel>[, <Inductance>]
Specifies the inductance value L in the "Shunt L, Shunt C" lumped element model for
differential match embedding.
In the query form, the <Inductance> parameter must be omitted.
Suffix:
<Ch>
.
Channel number
<LogPt>
Logical port number (balanced port)
<Cmp>
Must be omitted or set to 1
Parameters:
<CircuitModel>
SHLC
Currently only the "Shunt L, Shunt C" lumped element model is
supported
<Inductance>
Range:
-1 H to 1 H
Increment: 1 pH (1E-12 H)
*RST:
1 nH (1E-9 H)
Default unit: H
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:EMBedding<LogPt>:
PARameters:R<Cmp>
<CircuitModel>[, <Resistance>]
Specifies the resistance value R in the "Shunt L, Shunt C" lumped element model for
differential match embedding.
In the query form, the <Resistance> parameter must be omitted.
Suffix:
<Ch>
.
Channel number
<LogPt>
Logical port number (balanced port)
<Cmp>
Must be omitted or set to 1.
Parameters:
<CircuitModel>
SHLC
Currently only the "Shunt L, Shunt C" lumped element model is
supported
<Resistance>
Range:
-10 MΩ to 10 MΩ.
Increment: 1 mΩ (1E-3 Ω)
*RST:
0 Ω
Default unit: Ohm
Manual operation:
See
SCPI command reference
www.allice.de