GUI reference
R&S
®
ZND
436
User Manual 1173.9557.02 ─ 61
5.11.4.2
Active Power Cals dialog
The "Active Power Cals" dialog shows the power calibrations that are compatible with
the active channel configuration. It allows you to enable or disable the available power
calibrations.
Access:
CHANNEL – [CAL] > "Use Cal" > "Active Power Cals..."
All possible power calibrations can be performed from the
Port Overview
Shows all source ports together with their possible power calibrations. Unavailable cali-
brations (i.e. calibrations that have not been performed yet) are grayed out.
Note:
For a unidirectional R&S
ZND (see
on page 80) P2 cannot serve as source port. Hence only a
calibration of the measurement receiver can be performed at P2.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection:STATe
[SENSe<Ch>:]CORRection:POWer<PhyPt>[:STATe]
[SENSe<Ch>:]CORRection:PCAL
5.11.4.3
Calibration Manager dialog
The "Calibration Manager" dialog stores user correction data to the cal pool and
assigns stored correction data to channels.
See
Chapter 4.5.3, "Calibration pool"
on page 163 for background information.
Access:
CHANNEL – [CAL] > "Use Cal" > "Cal Manager..."
Cal softtool