GUI reference
R&S
®
ZND
394
User Manual 1173.9557.02 ─ 61
Figure 5-13: Calibration Unit wizard, step 2: Connections (>4 test ports and multiple port assign-
ments)
If multiple port assignments are required and
"Use Reduced Number of Through"
on page 543 is enabled, make sure that each port assignment contains the
.
Port Assignment (manual)
The "Connections" step allows you to configure the assignment between VNA ports
and calibration unit ports manually.
The test port connectors are automatically set according to the connector type of the
selected calibration unit port.
Detect Port Assignment
Starts a procedure by which the R&S
ZND (with a little help from the attached calibra-
tion unit) auto-detects the connected ports. The automatic assignment replaces the
configured one.
If auto-detection fails because of a high attenuation in the signal path, you can either
enter the port assignment manually or connect matching port numbers and select
"Default Port Assignment".
Note:
For a unidirectional R&S
on page 80) auto-detection at VNA port 2 does not work.
Remote command:
[SENSe:]CORRection:COLLect:AUTO:PORTs:CONNection?
Default Port Assignment
Restores the default port assignment.
Start
Proceeds to
.
If the configured port assignments are invalid, this action is disabled.
Cal softtool
www.allice.de