Concepts and features
R&S
®
ZND
211
User Manual 1173.9557.02 ─ 61
●
For a unidirectional R&S
ZND, only port 1 can be used for distance-to-fault mea-
surements.
●
A more flexible approach with additional configuration possibilities is available with
time domain option R&S
Chapter 4.7.5, "Time domain analysis"
4.7.7
Frequency resolution 1 mHz
Option R&S
ZND-K19
This software option improves the frequency resolution of a R&S
ZND to 1
mHz.
4.7.8
Eazy de-embedding based on IEEE 370
Option R&S
ZND-K210
EaZy deembedding (EZD) is an implementation of the IEEE algorithm for generating
deembedding files from a symmetrical 2x thru measurement.
Option R&S
ZND-K210 allows you to model a test fixture using the EZD fixture model-
ing tool and to deembed selected ports using the generated touchstone files.
The EZD fixture modeling proceeds in the following steps:
1. Measure one or more test coupons for the related fixture.
2. Run the EZD tool to calculate the Touchstone files modeling the test fixture.
The calculated
sNp
files can then be used to deembed the DUT at selected ports.
Optional extensions and accessories
www.allice.de
Allice Messtechnik GmbH