Instrument functions
R&S
®
NGU
51
User Manual 1179.2531.02 ─ 03
Figure 6-10: Select trigger dialog
3. Select the required trigger action.
● "Output On": If triggered, the selected output channel is turned on.
● "Output Off": If triggered, the selected output channel is turned off.
● "Gated": If triggered, the selected output channel is gated.
4. Select "Set" to configure the trigger action.
6.2.3.4
Output mode
1. Select the "Output Mode" menu item to set the output mode. See
The R&S
NGU displays the "Select Output Mode" dialog.
Figure 6-11: Select output mode dialog
2. Select the required output mode.
● Auto: Depending on the voltage across the output terminal, the R&S
NGU goes
into sink or source mode.
● Sink : The R&S
NGU goes to sink mode, current flows into the instrument. On
the display, sink mode is in operation if negative current in shown for NGU201.
However, if both voltage and current are shown as opposite signs (e.g. nega-
tive voltage and positive current), it indicates that NGU401 is operated in the
sink mode.
Activating the channel output