Alia ALIAVTX AVF7000 Series Скачать руководство пользователя страница 1

AVF7000 Operation Manual   

                                                                                                          

OP7000.1.1.7.R12ENG 

 
 
 

 

Vortex Flowmeter 

 

AVF7000 Series 

 

Operation Manual 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 

 
 
 

ALIA TECHNOLOGY LLC 

URL: http://www.alia-inc.com

633 W. 5th Street, 26th Floor, Los Angeles, CA 90071, USA 

E-mail: [email protected]

TEL : +1 - 213 - 533 - 4139         FAX : +1 - 213 - 223 - 2317 

OP7000.1.1.7.R12ENG

A L I A V T X

®

 

Содержание ALIAVTX AVF7000 Series

Страница 1: ...G 1 Vortex Flowmeter AVF7000 Series Operation Manual ALIA TECHNOLOGY LLC URL http www alia inc com 633 W 5th Street 26th Floor Los Angeles CA 90071 USA E mail alia alia inc com TEL 1 213 533 4139 FAX...

Страница 2: ...and 4 20mA 2 wire or Scale pulse output 9 7 4 With Temp Pressure Compensation and RS485 Communication 10 7 5 Separation board 11 8 Panel Display 12 9 Function 12 9 1 Button Function 12 9 2 Display Fu...

Страница 3: ...emical engineering pharmacy paper making metallurgy electricity environmental protection and food 3 Working Principle ALIA AVF7000 is newly stress type flowmeter that s based on von Karman effect It s...

Страница 4: ...atability 0 2 Load 1000 000 Process Connection Flange Wafer Communication HART or RS485 Flange Type JIS 10K JIS 20K JIS 40K Data Storage Operation Parameter Totalizer Figures are stored by EEPROM ANSI...

Страница 5: ...500 500 298 300 12 750 785 750 450 550 550 350 14 805 840 805 500 600 600 400 16 861 895 861 550 650 650 450 18 910 945 910 600 700 700 500 20 965 998 965 650 750 750 Wafer Type Flange Type Flange Ty...

Страница 6: ...t concerns measuring accuracy Please leave more space of straight pipe in upstream and downstream as best as you can If not permitted please install as 2 3 of total straight pipe length on the upstrea...

Страница 7: ...pictures all you need to do is to ground the displayer And the sensor grounding is unnecessary Grounding place is like stairs and railing To define whether vortex is well grounded initially user need...

Страница 8: ...onnected to the housing of amplifier 7 1 With 4 20mA 2 wire or Scale pulse but without Temp Pressure Compensation 4 20mA 2 wire output 4 20mA and Scale Pulse Output 7 2 With RS485 Communication but wi...

Страница 9: ...tput Sensor Input 4 20 mA Output Pulse Output Temperature Pressure mA 12 32VDC Power Supply A 4 20mA 2 wire 12 32 VDC Power Supply 1 5KOhm Pulse Output A T T V V I I A A Current Outout Voltage Input P...

Страница 10: ...emp Pressure Compensation and RS485 Communication Sensor Input 12 32 VDC Power supply RS485 Communication Temperature Pressure 12 32VDC Power Supply A B A B RS485 RS485 T T V V I I A A Current Outout...

Страница 11: ...rd Note The terminals S1 S2 are signal ports of vortex sensor Terminal Cable color Sensor PI red PI blue PV black PV yellow Pressure sensor T white TA red TB red Temperature sensor S1 orange S2 orange...

Страница 12: ...Frequency Density Pressure Temperature Current Percentage Temperature Pressure Other display If enable pressure auto acquisition and pressure signal is abnormal the pressure value in Temp Pressure int...

Страница 13: ...dium is gas or steam it s 80000 Amplification adjustment method Magnification can be divined into two parts amplification factor and magnification gain Magnification factor is in Max AMP while magnifi...

Страница 14: ...ation Chart SET SET Contrast Protection Meter Size Flow Mode Unit Range 100 Density Kg m3 Gauge Pre Kpa PV Cutoff Damping S Min Alarm Max Alarm SET SET 10 5 m3 h E 20 0 m3 P 12 5Kpa T 20 0 Disp Point...

Страница 15: ...assword Pulse Factor Unit Pulse Factor SET SET CODE SET SET 4mA Trim 20mA Trim CODE SET SET SET SET K Factor Trim Y1 K Factor Trim F1 K Factor Trim Y2 K Factor Trim F2 K Factor Trim Y3 K Factor Trim F...

Страница 16: ...sword AMP Channel CODE SET SET SET SET Work Mode CODE SET SET SET SET Temp Low Trim Temp High Trim Temp Measure Pre Low Trim Pressure Measure Pre High Trim Pre Cutoff Set Pre Bias Modbus Addr Modbus B...

Страница 17: ...m 200 000 Min Alarm 100 000 Alarm High Limit Set by and keys for instance 200 it means alarm high limit is 200 Full scale If displayed value is more than the high limit the current output is 22mA This...

Страница 18: ...set totalizer Press to enter next setting interface Total Overflow 0 000 Totalizer Overrange Cycle This interface is read only When flow total is over 99999999 the overrange cycle will be 1 And then f...

Страница 19: ...by and keys Enter advanced settings by password Press to return parameter protection interface Press SET SET to back to normal display interface SET SET 10 5 m3 h E 20 0 m3 K Factor 9100 00 K Factor R...

Страница 20: ...ssary please contact ALIA Press to enter next setting interface Low Flow Limit 35 000m3 h Low Flow Limit Set by and keys Theoretical min flow rate this parameter does not affect the measurement for re...

Страница 21: ...l the ammeter displays 4mA Each time you press 0 016mA will be changed After calibration press long for 3s to save calibration or press SET SET to exit without saving calibration After calibrating 4mA...

Страница 22: ...he initial value of each point is 1 0000 E g When you find the flow rate is not displaying accurate you can check corresponding flow rate of different velocities in normal display interface and press...

Страница 23: ...ease until value 2500 00 flashes after this press for 3s again until value 2500 Ohm stops flashing then release Temp High point calibration is finished Press to enter next setting interface Temp High...

Страница 24: ...ng interface Pre Cutoff 0 000kpa SET SET Set Pre Bias 0 00kpa 10 5 m3 h 20 0 m3 SET SET Pressure Shift Set by and keys Unit KPa If pressure is auto measuring and displaying pressure is inaccurate you...

Страница 25: ...topbit 1 Communication address 01 Password Press and to enter password 00090 Hold for 3s to enter MODBUS settings Code 00090 Communication Address Press and to set range 1 247 Default address 1 Press...

Страница 26: ...f necessary If measuring medium is steam please make sure actual temperature pressure parameters are conformed with those inside circuit board Circuit board or sensor is faulted Please send them back...

Отзывы: