Calibration
46
16.5.1 1V to 2V Range
Shift+Serv|Password|Calibration|Voltage Channel | 1-2V Range
1)
From the Calibration menu, select Voltage Channel, then 1 - 2V range.
2)
Adjust the Cal source for 2.000 volts
100mV as indicated on the DVM.
3)
Press the Up/Down arrows on the CRT until the voltage displayed on
the CRT matches the reading on the DVM.
4)
Record the actual input and the CRT reading on the Calibration form.
5)
Press F2 Accept when finished.
6)
From the Calibration menu, select Voltage Channel, 1 - 2V range.
7)
Linearity Check: Adjust the Cal source to 1.000V
100mV on the DVM
and record the actual input and CRT reading.
8)
Linearity: Adjust the Cal source to 2.350V
100mV on the DVM and
record the actual input and CRT reading.
9)
Press F3 Cancel.
16.5.2 4V to 6V Range
Shift+Serv|Password|Calibration|Voltage Channel | 4-6V Range
1)
From the Calibration menu, select Voltage Channel, then 4 - 6V range.
2)
Adjust the Cal source to 5.000 volts
100mV on the DVM.
3)
Press the Up/Down arrows until the voltage on the CRT matches the
DVM.
4)
Record the actual input and the CRT reading on the form.
5)
Press F2 Accept.
6)
From the Calibration menu, select Voltage Channel, 4 - 6V range.
7)
Linearity: Adjust the Cal source to 4.000V
100mV on the DVM and
record the actual input and CRT reading.
8)
Linearity: Adjust the Cal source to 7.000V
100mV on the DVM and
record the actual input and CRT reading.
9)
Press F3 Cancel.