background image

alarm will trip only if all sensor readings are 85°F or above 
for a continuous 30-minute period. 

7.

 

Repeat steps 2 through 6 for the other sensor if desired. 

8.

 

Choose how long the logger should maintain a sensor alarm 
once it has tripped. Select “Host has relaunched logger” if 
you want the alarm to remain visible on the LCD until the 
next time you relaunch the logger. Select “Sensor reading 
within limits” if you want the alarm to clear once the sensor 
reading returns to the normal range between the high and 
low alarm limits. Select “Cleared with button press” if you 
want the alarm to remain on until you press the Alarm/Stats 
button on the logger. 

9.

 

Click OK to save alarm settings. 

Notes: 

 

Once the logger is launched, alarms will trip as 
determined by these settings. Logger alarms will display 
on the LCD screen. Note that the alarm limits are only 
checked when the logger’s LCD screen refreshes every 15 
seconds. 

 

The actual values for the high and low alarm limits are set 
to the closest value supported by the logger. For 
example, the closest value to 85°F that the UX100 series 
logger can record is 84.990°F and the closest value to 
32°F is 32.043°F. In addition, alarms can trip or clear 
when the sensor reading is within the logger 
specifications of 0.02°C resolution. This means the value 
that triggers the alarm may differ slightly than the value 
entered. For example, if the High Alarm is set to 
75.999°F, the alarm can trip when the sensor reading is 
75.994°F (which is within the 0.02°C resolution). 

 

When you read out the logger, high and low alarm levels 
will be displayed on the plot along with “Chan <#> Alarm 
Tripped” and “Chan <#> Alarm Cleared” events showing 
when the sensor alarm tripped and cleared. The “Chan 
<#> Alarm Cleared” event contains the value that was 
furthest out of range for the sensor before the alarm 
cleared (see the Points table for the actual value). 

Burst Logging 

Burst logging is a logging mode that allows you to set up more 
frequent logging when a specified condition is met. For 
example, let’s say the logger is recording data at a 5-minute 
logging interval and burst logging is configured to log every 10 
seconds when the temperature goes above 85°F (the high limit) 
or falls below 32°F (the low limit). This means the logger will 
record data every 5 minutes as long as the temperature 
remains between 85°F and 32°F. Once the temperature reaches 
90°F, for example, the logger will switch to the faster logging 
rate and record data every 10 seconds until the temperature 
falls back below the high limit (or 85°F in this case). At that 
time, logging then resumes every 5 minutes at the normal 
logging interval. Similarly, if the temperature falls to 30°F, for 
example, then the logger would switch to burst logging mode 
again and record data every 10 seconds. Once the temperature 
rises back to 32°F, the logger will then return to normal mode, 
logging every 5 minutes. 

To set up burst logging: 

1.

 

Select Burst for Logging Mode in the Launch Logger 
window. If Burst has already been configured for this 
logger, click the Edit button in the Launch Logger window. 

Note that you cannot set up alarms with burst logging. You 
also cannot select the Stop Logging option “Never (wrap 
when full)” if burst logging is configured. 

2.

 

In the Burst Logging window, select a sensor. In the 
example, the temperature sensor was selected. 

3.

 

Select the High Limit checkbox if you want to set up a 
condition in which burst logging will occur when the sensor 
reading rises above the high limit value. Type in the value or 
drag the red upper slider. 

4.

 

Select the Low Limit checkbox if you want to set up a 
condition in which burst logging will occur when the sensor 
reading falls below the low limit value. Type in the value or 
drag the blue lower slider. 

 

5.

 

Repeat steps 2 through 4 for the other sensor, if desired. 

6.

 

Set the burst logging interval, which must be less than the 
logging interval. Select either a preset burst logging interval 
or select Custom and enter your own interval. Keep in mind 
that the more frequent the burst logging rate, the greater 
the impact on battery life and the shorter the logging 
duration. 

7.

 

Click OK when done. This will return you to the Launch 
Logger window. Click the Edit button next to Logging Mode 
in the Launch Logger window to make additional changes. 

Notes: 

 

Once the logger is launched, the high and low burst 
logging limits are only checked when the logger’s 
LCD screen refreshes once every 15 seconds. Therefore, 
if you set the logging interval to less than 15 seconds and 
the sensor reading falls outside the limits, the burst 
logging will not begin until the next 15-second refresh 
cycle. 

 

If high and/or low limits have been configured for more 
than one sensor, then burst logging will begin when any 
high or low condition goes out of range. Burst logging will 
not end until all conditions on all sensors are back within 
normal range. 

 

The actual values for the burst logging limits are set to 
the closest value supported by the logger. For example, 
the closest value to 85°F that the logger can record is 
84.990°F and the closest value to 32°F is 32.043°F. 

Содержание Hobo UX100-003

Страница 1: ...0 to 95 RH non condensing Launch Readout 0 to 50 C 32 to 122 F per USB specification Logging Rate 1 second to 18 hours 12 minutes 15 seconds Logging Modes Fixed interval normal burst or statistics Mem...

Страница 2: ...logger is equipped with an LCD screen that displays details about the current status This example shows all symbols illuminated on the LCD screen followed by definitions of each symbol in the table b...

Страница 3: ...Select Turn LCD off when setting up the logger as described in the next section When this option is enabled you can still temporarily view the LCD screen by pushing the Start Stop button for 1 second...

Страница 4: ...and restart the logger with the Start Stop button 9 Choose whether to keep the LCD on or off By default the LCD will always remain on while logging If you select the Turn LCD off checkbox the LCD wil...

Страница 5: ...ecord data every 10 seconds until the temperature falls back below the high limit or 85 F in this case At that time logging then resumes every 5 minutes at the normal logging interval Similarly if the...

Страница 6: ...H data logged every 5 minutes plus eight maximum minimum average and standard deviation series four for temperature and four for RH with values calculated and logged every 5 minutes based on the 30 se...

Страница 7: ...et HUM UPS 500 To replace the sensor 1 Use a screwdriver to pop out the louvered door in the logger case 2 Use needle nose pliers to remove the RH sensor from the board inside the logger case and disc...

Страница 8: ...fire Do not expose the contents of the battery to water Dispose of the battery according to local regulations for lithium batteries HOBOware provides the option of recording the current battery volta...

Отзывы: