ALAT UJI C.A 1864 Скачать руководство пользователя страница 1

     

 
 

C.A 1864 
C.A 1866

 

 

 

 

 

§

 

Thermomètres infrarouge 

§

 

Infrared thermometers 

§

 

Infrarot - thermometer 

§

 

Termometri a infrarossi 

§

 

Termómetros infrarrojo 

User’s manual 

E N G L I S H  

 

 

 

Office: Jl. Radin Inten II No. 62 Duren Sawit, Jakarta 13440 - Indonesia
Workshop: Jl. Pahlawan Revolusi No. 22B, Jakarta 13430 - Indonesia
Phone: 021-8690 6777 (Hunting)

Fax: 021-8690 6777

Mobile: +62 816 1740 8925

Fax: 021-8690 6771

Содержание C.A 1864

Страница 1: ...mometri a infrarossi Term metros infrarrojo User s manual E N G L I S H Office Jl Radin Inten II No 62 Duren Sawit Jakarta 13440 Indonesia Workshop Jl Pahlawan Revolusi No 22B Jakarta 13430 Indonesia...

Страница 2: ...ecautions detailed MEANING OF THE SYMBOL Sorting of wastes required for the recycling of electrical and electronic equipment in the European Union In accordance with directive WEEE 2002 96 EC this equ...

Страница 3: ...temperature is 65 C 150 F If the thermometer is subjected to great variations in ambient temperature wait for 30 minutes after this becomes stable again before continuing with measurement so as to at...

Страница 4: ...MENT PARAMETERS 24 4 5 ERROR MESSAGES 26 5 SPECIFICATIONS 27 5 1 SPECIFICATIONS 27 5 2 CLIMATICS CONDITIONS 27 5 3 CONSTRUCTION SPECIFICATIONS 27 5 4 CONFORMITY TO INTERNATIONALS STANDARDS 27 6 WARRAN...

Страница 5: ...3 15 C give out infrared energy This radiated energy is emitted in all directions at the speed of light The instrument s lens picks up this energy and focuses it on an infrared detector This detector...

Страница 6: ...s of emissivity values hight and low value of the alarm 5 r increase in increments of emissivity values hight and low value of the alarm 6 MODE choice of measurement parameters through successive pres...

Страница 7: ...the LASER sight and backlighting 4 Aim the thermometer whilst pressing the trigger using the LASER sight or otherwise at the target to be measured see 4 2 and 4 3 Note The instrument memorises the la...

Страница 8: ...he probes measurement field at a given distance from the target see figure below In the case of a measurement point of small dimensions it is important to bring the probe to a short enough distance fr...

Страница 9: ...djustment respectively The setting is held in memory after the instrument is switched off and used during all successive measurements whatever the parameter used MAX see fig to right When the trigger...

Страница 10: ...asured as of the start of measurement The measurement set up is held in memory after the instrument is switched off but not the last temperature difference value recorded since the instrument takes in...

Страница 11: ...se slide the bottom switch SET ALARM right To set values for the High Alarm HAL Low Alarm LAL firstly active the display by pulling the trigger or press MODE then press MODE until the appropriate code...

Страница 12: ...1s Spectral reponse 8 to 14 m Emissivity adjustable from 0 1 to 1 Screen lighting adjustable by the user Measurement functions normale MAX MIN average value AVG difference between starting value DIF c...

Страница 13: ...y type 9V is required Open battery cover then take out the battery from instrument and replace with a new 9 Volt battery Place the battery cover back 7 2 WORKING TEST To check the good working order o...

Отзывы: