BDDF00144 Issue 1
20/40
Alaris™ GW 800 Volumetric Pumps
Basic Features
Changing the Infusion Set
1 . Press
d
to put the Pump on
G
.
2 . Close in-line clamp and ensure the IV access to the patient is isolated .
3 . Disconnect the Infusion set from the patient .
4 . Open Pump door and remove Infusion set from the Pump and discard the set and fluid container according to hospital protocol .
5 . Place new Infusion set into Pump, see 'Loading the Infusion Set' .
6 . Squeeze the drip chamber approximately half full or up to fill line (if the drip chamber is marked) with fluid .
7 . Prime the set manually .
8 . Restart infusion, see 'Getting Started' .
w
When changing the Infusion set or the fluid container use aseptic technique according to hospital protocol. It
is recommended that Infusion sets are changed in accordance with the Directions for Use. Carefully read the
Directions For Use supplied with the Infusion set prior to use. The set change interval is 24 hours.
Changing the Fluid Container
1 . Press
d
to put the Pump on
G
.
2 . Remove bag spike on Infusion set from empty / used container . Discard empty / used container according to hospital protocol .
3 . Insert spike into new container .
4 . Squeeze the drip chamber approximately half full or up to fill line (if the drip chamber is marked) with fluid .
5 . Restart infusion, see 'Getting Started' .
w
When changing the Infusion set or the fluid container use aseptic technique according to hospital protocol. It
is recommended that Infusion sets are changed in accordance with the Directions for Use. Carefully read the
Directions For Use supplied with the Infusion set prior to use.
SmartSite™ Needle-Free System Instructions
SmartSite Needle-Free Valve is designed to permit safe gravity flow and automated flow, injection and aspiration of fluids without the
use of needles by utilising Luer lock and Luer slip connectors .
w
Precautions:
Discard if packaging is not intact or protector caps are unattached.
If SmartSite Needle-Free Valve is accessed by a needle in an emergency the valve will be damaged causing leakage.
Replace SmartSite Needle-Free Valve immediately.
SmartSite Needle-Free Valve contraindicated for blunt cannula system.
Do not leave slip Luer syringes unattended.
DIRECTIONS - Use Aseptic Technique
1 . Prior to every access, swab top of SmartSite Needle-Free Valve port with 70% Isopropyl alcohol (1-2 seconds) and allow to dry
(approximately 30 seconds) .
NOTE:
Dry time is dependent on temperature, humidity, ventilation of the area .
2 . Prime valve port . If applicable, attach syringe to SmartSite Needle-Free Valve port and aspirate miniscule air bubbles .
3 . When used with administration sets always refer to individual set directions for use as change interval may vary according to clinical
application (e .g . infusions of blood, blood products, and lipid emulsions) .
NOTE:
During use of Needle-Free Valve port, fluid may be observed between the housing and blue piston . This fluid does not enter
the fluid path and requires no action .
NOTE:
For product questions or needle-free valve educational materials, contact your BD representative . Consult facility protocols .
Consult other organisations that publish guidelines useful in developing facility protocols .