Maintenance
CRIUS II
Main power supply
6.7.3
Checking the fuses
The fuses are located in the electrical cabinet.
Fig. 6-7
Fuses
1
Open the corresponding doors
2
Check the switching function of the fuses by means of a voltmeter
6.7.4
Checking the over-current circuit breakers
The system contains over-current circuit breakers. They are located in the
electrical cabinet and in the heater supply unit.
Fig. 6-8
Over-current circuit breakers
A Switch
B Over-current value
C «Test» switch
1
Open the corresponding doors
2
Check the correct setting of the over-current value
•
The required value is 40 A
3
Activate the «Test» switch using a screw driver
•
The circuit breaker must switch off
4
Switch on the circuit breaker after the test procedure