![AIXTRON CRIUS II Скачать руководство пользователя страница 148](http://html1.mh-extra.com/html/aixtron/crius-ii/crius-ii_system-manual_2883093148.webp)
cr
ius_I
I_e
n_
00,
Ed
it
io
n
06/
201
0
148
CONFIDENTIAL
AIXTRON
- Dokumentation
Maintenance
System Manual
Cleaning procedures
CRIUS II
6
6.6.4
Stainless steel parts
Solutions
AIXTRON recommends a Sodium Hydroxyde solution on systems growing
Arsenides.
If grown layers are predominantly Phosphides AIXTRON recommends to use
a solution of orthophosphoric acid (30%) and hydrogen peroxide (30%) in DI
water.
H
2
O : NaOH
10 : 1 or
H
2
O : H
3
PO
4
: H
2
O
2
8 : 1 : 1