![AIXTRON CRIUS II Скачать руководство пользователя страница 136](http://html1.mh-extra.com/html/aixtron/crius-ii/crius-ii_system-manual_2883093136.webp)
cr
ius_I
I_e
n_
00,
Ed
it
io
n
06/
201
0
136
CONFIDENTIAL
AIXTRON
- Dokumentation
Maintenance
System Manual
Auxiliary devices and used materials
CRIUS II
6
6.2.3
Solid waste materials
The following list specifies parts that become solid waste as a result of opera-
tion, maintenance and servicing of the CRIUS II. The disposal of these parts
must be regulated by the end user according to the local safety regulations.
•
Gas inlet filters
•
Particle filter
•
Glove box filters
•
Vacuum cleaner filters
•
Vacuum wand filters
•
Circulation filters
6.2.4
Operating materials
•
Cleaning agents
•
Lint-free cloth
•
Isopropanol (2-propanol)
NOTE:
Parts of the CRIUS II to be decommissioned or replaced may be contam-
inated.
The RMA (Return Material Authorization) procedure exactly describes
the handling of contaminated parts.
WARNING!
Unsuitable solvents.
Acetone, methanol and ethanol are unsuitable solvents. These chemi-
cals attack seals made of nitrile and fluorocarbon rubber (e.g. Perbunan
and Viton). In addition, methanol is toxic.