3.0 Operation

3.1 

GENERAL 

If the Workhorse Oxygen Concentrator is supplied with a cover, make certain the 

fan cord is connected to the fan on the cover, before reinstalling the cover. The 

length of the cord provides for cover removal without disconnecting the cord from 

the fan.

To start the Workhorse Oxygen Concentrator, connect the unit to a grounded power 

source rated for the voltage and current requirements stated on the label on the 

unit. Push the toggle switch on the front panel to the up position. [The green LED 

on the front panel of the unit will light, indicating that power is applied to the system, 

and the red LED will light indicating low oxygen concentration.]

Set  the  outlet  oxygen  flow  to  4  LPM  (8  SCFH)  or  less.  [When  the  oxygen 

concentration achieves 70%, the red LED will go out and a yellow LED will light. 

When the concentration is greater than 85%, the yellow LED will go out.]

[The visual LED alarms also correspond to two alarm outputs on the nine pin D-sub 

connector located under the LEDs. These outputs are Form A one amp dry contacts. 

They are normally open and then close during an alarm. Only one contact is closed 

at any time during an alarm. If the oxygen concentration falls below 85%, the yellow 

LED will light and pins 4 and 5 will close. If the concentration falls below 70%, these 

contacts will open and the yellow LED will turn off. The red LED will then light and 

pins 2 and 3 will close.]

[Pins 1 and 6 can be used to measure the concentration level to within ±2% of 

concentration. This is read by placing the negative lead of a digital voltmeter on pin 

1 and the positive lead on pin 6. A 0-1 volt output is displayed with one volt relating 

to 100% and zero volts relating to 0%. Audible and visual alarms, automatic phone 

dialers and digital oxygen purity displays that directly interface with the nine pin 

connector are available as accessories. Contact AirSep for additional information.] 

[ ] Applies to models with optional oxygen monitor.

The flow meter installed on the Workhorse Concentrator is set to 

read accurately when the discharge is to atmospheric pressure. If 

the actual discharge pressure is substantially above atmospheric 

pressure, the reading can be adjusted to determine the precise 

flow rate, according to the following formula:
(adjusted flow) = (measured flow) x √

oxygen pr 14.7 

 

 

 

 

 

          14.7

Please contact AirSep Technical Support if additional assistance is required.

Do not allow the oxygen product to vent freely. Do not exceed rated capacity.

Содержание SeQual Workhorse Series

Страница 1: ...INSTRUCTION MANUAL WORKHORSE SERIES SeQual Workhorse 8 PSA Oxygen Concentrator System ...

Страница 2: ...6 2 0 Installation 2 1 OXYGEN HOOK UP 7 2 2 COMPRESSOR INLET FILTER 8 3 0 Operation 3 1 GENERAL 9 4 0 Maintenance 4 1 GENERAL 11 5 0 Troubleshooting 5 1 GENERAL 13 6 0 Technical Data 6 1 SPECIFICATIONS 15 6 2 DRAWINGS AND SCHEMATICS 16 7 0 Service Parts 7 1 SERVICE PARTS 17 7 2 REPLACEMENT TUBING 18 7 3 TECHNICAL SUPPORT 18 ...

Страница 3: ...List of Figures and Tables Table 6 1 Workhorse 8 Specifications 15 Figure 6 1 Mechanical Outline Drawing Workhorse 8 16 Table 7 1 Workhorse 8 Service Parts 17 Table 7 2 Replacement Tubing 18 ...

Страница 4: ......

Страница 5: ...ily ignited and burn rapidly in high concentrations of oxygen For safety concerns all possible sources of ignition must be kept away from the oxygen concentrator and preferably out of the room in which it is being used Observe strict cleanliness procedures when fabricating and connecting the oxygen piping It is imperative that oxygen systems be properly cleaned and inspected to insure that no comb...

Страница 6: ...anual AirSep Corporation 1 3 PATENTS TRADEMARKS AND COPYRIGHT The Workhorse Oxygen Concentrator is protected by the following US Patents 4 925 464 5 112 367 5 114 441 5 268 021 5 366 541 5 593 478 Re 35 099 Other patents pending ...

Страница 7: ...on the label on the unit There are packaging materials that must be removed from inside the Workhorse enclosure before operation Do not attempt to operate the Workhorse Oxygen Concentrator without removing these packaging materials as severe damage will result The location of the Workhorse Oxygen Concentrator must be well ventilated to prevent suffocation of personnel Refer to the recommendations ...

Страница 8: ...en systems be properly cleaned and inspected to insure that no combustible materials remain in the system If you are not familiar with oxygen cleaning procedures refer to the Compressed Gas Association documents G 4 1 Cleaning Equipment for Oxygen Service and G 4 4 Industrial Practices for Gaseous Oxygen Transmission and Distribution Piping Systems Do not allow the free flow of oxygen from the Wor...

Страница 9: ...nly one contact is closed at any time during an alarm If the oxygen concentration falls below 85 the yellow LED will light and pins 4 and 5 will close If the concentration falls below 70 these contacts will open and the yellow LED will turn off The red LED will then light and pins 2 and 3 will close Pins 1 and 6 can be used to measure the concentration level to within 2 of concentration This is re...

Страница 10: ......

Страница 11: ... approximately every 4 000 hours of operation Filter change frequency is dependent on environmental conditions and may vary Compressors require a rebuild after 5 000 to 12 000 hours of operation depending on environmental conditions See the Service Parts section for information on obtaining filters and compressor rebuild kits ...

Страница 12: ......

Страница 13: ... R E G N A H C X E S S E L L I O R O S S E R P M O C G U L P E L B A T S U J D A W O L F R E T E M A P E H R E T L I F L A N O I T P O E L U D O M N E G Y X O D R A O B N O F T A E L U D O M E R U S S E R P F E I L E R E V L A V R E Z Y L A N A N E G Y X O T U O O2 In our normal operation of an ATF the input and output pressures should be within about 2 psig of what they were when the unit was new...

Страница 14: ...n it shipped Other symptoms If the pressures are fluctuating more than about 2 psig around the median it could indicate an internal rupture or a previous incidence of the motor not turning If the pressure fluctuates more than 3 psig on the air input this might indicate that one or more of the flow controlling orifices inside is compromised That would mean when the cycle pressurizes these beds the ...

Страница 15: ...rom weather elements and remains between 40 F 4 C and 110ºF 43ºC inside room or operating enclosure 20ºF 29ºC and 170ºF 77ºC in storage Control Power Requirements Single Phase 120 V 10 60 Hz 2 8 A 230 V 10 50 Hz 1 8 A 230 V 10 60 Hz 1 8 A Typical Power Consumption 325 W Compressed Air Pressure Relief Valve setting 35 psig 10 Environment Barometric Pressure Range 28 31 of Hg Ambient Oxygen Concentr...

Страница 16: ...oration 6 2 DRAWINGS AND SCHEMATICS 14 36 FILTER SILENCER FLOWMETER HEAT EXCHANGER COMPRESSOR ATF POWER CORD POWER SWITCH FRONT VIEW SIDE VIEW 17 43 18 46 O2 MONITOR HOUR METER PRESSURE GAUGE 14 36 Figure 6 1 Mechanical Outline Drawing Workhorse 8 ...

Страница 17: ... 60 Hz 1265V SEQ 1 ATF Module 220 V 50 Hz 7465V SEQ 1 Compressor 120 V 60 Hz 7329 SEQ 1 Compressor Rebuild Kit for 120 V 60 Hz or 220 V 50 Hz 7366 SEQ 1 Compressor 220 V 50 Hz 7330 SEQ 1 Compressor 230 V 60 Hz 1619 SEQ 1 Compressor Rebuild Kit for 220 V 60 Hz 3197 SEQ 1 Cylinder Walls aka Cylinder Sleeves for 220 V 60 Hz compressor if needed 3411 SEQ 2 Compressor Inlet Filter 3308 SEQ 1 Compressor...

Страница 18: ...r assistance in troubleshooting or repairing the unit or to order replacement parts contact the AirSep Commercial Products Service Department by telephone In the USA or Canada call 1 800 320 0303 Outside the USA or Canada call 716 691 0202 Send fax inquiries anytime to 716 564 2717 Address written inquiries to AirSep Corporation 260 Creekside Drive Buffalo NY 14228 2075 USA Attention Commercial Pr...

Страница 19: ......

Страница 20: ...ty descriptions specifications and or processes to its products at any time without prior notice and with no further obligation or consequence Any rights expressly stated herein are reserved by us as applicable AirSep Corporation 260 Creekside Drive Buffalo NY 14228 2075 U S A Tel 716 691 0202 Fax 716 691 1255 info airsep com www airsepcpd com MADE IN USA ...

Отзывы: