Data Export 

 
By  the  USB  port  of  the  device,  the  movement  data  can  be  transferred  to  a  PC.  To  do  this 
export,  the  tuned  on  -ADS  intelligent  must  be  connected  to  the  USB  port  of  a  PC.  The  hard-
drive  of the  ADS intelligent  will  be detected by the  PC. 
 

12.1

 

File format and using the data 

 
The  data  will  be  saved  in  "*.CSV"  format  and  can  be  opened  at  the  PC  by  Excel  of  Calc.  One 
file  per day will  be saved with  the name  "year  month  day.csv"   
 
Example:  a  data  file  with  the  name  "20160723.csv"  has  been  saved  and  recorded  at  23rd    of 
July  2016.  
 

 

 

Remark 1: it is only possible to do read out the data. Data onto the The hard-drive of the ADS 
intelligent cannot be modified, changed or saved by the PC.  
 
Remark  2:  while  the  ADS  intelligent  is  connected  to  the  PC,  no  pressure  data  will  be 
recorded!   

Содержание ADS Intelligent

Страница 1: ...5 7 Turning on the device 7 8 Screen the beginning after turning on the ADS intelligent 8 9 Settings 10 9 1 Weight setting 10 9 2 Alarms and warnings 11 9 2 1 PMD and alternating change pressure cycle...

Страница 2: ...he corresponding mattress can be used till pressure ulcers gradation 4 2 Operation the device Connecting the device to the power supply The ADS intelligent must connect to a power strip of 230 V 50Hz...

Страница 3: ...ect the mattress and the bed using the top straps Otherwise there is danger that the maneuvarability of the bed and even the functionality of bed will be limited In addition there is a danger of damag...

Страница 4: ...em is operational The patient may not guaranteed bed sores by pressure relief at the filling process and missing volume fill early The ADS intelligent features a log system that duration the error cod...

Страница 5: ...tests for electrical safety for electrical devices Annual maintenance is recommended because the ADS intelligent is a high quality medical device The product life cycle can be extended in case of regu...

Страница 6: ...Page 6 of 23 USER MANUAL ADS intelligent Softwareversion 0 95...

Страница 7: ...USER MANUAL ADS intelligent Softwareversion 0 95 7 Turning on the device After driving the main switch in position I the AirMed PLUS Logo and software version of the device is shown for approximately...

Страница 8: ...the given value pressure inside the cell is lower than the given value 2 This symbol shows that choosing patient s weight class 3 This symbol shows to enter the settings 4 Alarm signal It will be show...

Страница 9: ...ing please check chapter 9 2 Remark one the patient movement monitor can only be used in static therapeutic mode because the mode of alternating change pressure cycle the movement detection is not pos...

Страница 10: ...enu 9 1 Weight setting In this menu you can adjust the pressure inside the cells according to the patient s weight Press for increasing and press for decreasing There are 10 possible weight steps avai...

Страница 11: ...time no detected movement above the necessary movement level After every detected movement about the necessary level the counter will be reset and the movement detection starts from zero minutes agai...

Страница 12: ...Warnings In opposite to the alarm warnings will be only shown visually and not acoustically The number three 3 warnings can be configured also as acoustically or nurse call system Acoustic if enabled...

Страница 13: ...adjusted 9 3 2 Logic nurse call by this button the nurse call logic hardware relais can be permanently turned one or off If the nurse call logic is deactivated the relais will be activated in the case...

Страница 14: ...d Time 9 5 Date setup Please see chapter 10 Set up Date and Time 9 6 Touch calibration by using this function the touch screen will be calibrated The operator needs to the crosses in the display withi...

Страница 15: ...Page 15 of 23 USER MANUAL ADS intelligent Softwareversion 0 95 Inside this submenu the operator can select one language...

Страница 16: ...mode 9 8 1 Bedding mode In this mode the mattress will be inflated with full pressure for maximum 20 minutes After 20 minutes the system turns into the previous mode automatically again 9 8 2 head en...

Страница 17: ...s shown in the display and in the schematic mattress is marked by a small red bar above the motion monitor graph 1 By touching the time graph the operator can scroll horizontally from early to late in...

Страница 18: ...e current figure which is changed is highlighted in black colour 2 By pressing this button 2 the new settings will be saved 3 By pressing this button the operator will leave this submenu without savin...

Страница 19: ...patient monitoring a without redistributing any movement of patient please see chapter Error Reference source not found for details AND b Alternating cycle time mode is enabled One cycle of alternati...

Страница 20: ...le format and using the data The data will be saved in CSV format and can be opened at the PC by Excel of Calc One file per day will be saved with the name year month day csv Example a data file with...

Страница 21: ...Alarme sind m glich Warnings Priority Pump up 1 The mattress is still into pump up procedure and it is not allowed to place the patient onto the mattress If the pressure inside each cell is sufficient...

Страница 22: ...following medical device Product name ADS intelligent Meets the requirements of the European directive for medical products 93 42 EWG and 2007 47 EC The ADS clinic is consistent with risk class 1 bas...

Страница 23: ...USER MANUAL ADS intelligent Softwareversion 0 95 manufacturer of ADS intelligent AirMed PLUS GmbH Harpener Feld 34 D 44805 Bochum Germany Tel 0049 0 234 91773320 Fax 0049 0 234 91773322 Email service...

Отзывы: