USER MANUAL 

ADS intelligent 

 

 

 

 

Softwareversion: 

0.95 

 
 
 
 

 

10

 

Set up Date and Time   

 

 

 
 

1.

 

The  current  time  will  be  shown.  If  the  operator  is  changing  the  time,  the  current  figure 

which  is  changed  is  highlighted  in  black  colour.   

 
2.

 

By  pressing  this  button  (2) the new  settings  will  be saved. 

 

3.

 

By  pressing  this  button  the  operator will  leave  this  submenu  without  saving.   

 

4.

 

By  pressing  the  key-numbers  the  operator can change  the time  setting. 

 
To change  the  date the  operator can do it  like  the  same  when  changing  the  time. 

1

2. 

3. 

4

Содержание ADS Intelligent

Страница 1: ...5 7 Turning on the device 7 8 Screen the beginning after turning on the ADS intelligent 8 9 Settings 10 9 1 Weight setting 10 9 2 Alarms and warnings 11 9 2 1 PMD and alternating change pressure cycle...

Страница 2: ...he corresponding mattress can be used till pressure ulcers gradation 4 2 Operation the device Connecting the device to the power supply The ADS intelligent must connect to a power strip of 230 V 50Hz...

Страница 3: ...ect the mattress and the bed using the top straps Otherwise there is danger that the maneuvarability of the bed and even the functionality of bed will be limited In addition there is a danger of damag...

Страница 4: ...em is operational The patient may not guaranteed bed sores by pressure relief at the filling process and missing volume fill early The ADS intelligent features a log system that duration the error cod...

Страница 5: ...tests for electrical safety for electrical devices Annual maintenance is recommended because the ADS intelligent is a high quality medical device The product life cycle can be extended in case of regu...

Страница 6: ...Page 6 of 23 USER MANUAL ADS intelligent Softwareversion 0 95...

Страница 7: ...USER MANUAL ADS intelligent Softwareversion 0 95 7 Turning on the device After driving the main switch in position I the AirMed PLUS Logo and software version of the device is shown for approximately...

Страница 8: ...the given value pressure inside the cell is lower than the given value 2 This symbol shows that choosing patient s weight class 3 This symbol shows to enter the settings 4 Alarm signal It will be show...

Страница 9: ...ing please check chapter 9 2 Remark one the patient movement monitor can only be used in static therapeutic mode because the mode of alternating change pressure cycle the movement detection is not pos...

Страница 10: ...enu 9 1 Weight setting In this menu you can adjust the pressure inside the cells according to the patient s weight Press for increasing and press for decreasing There are 10 possible weight steps avai...

Страница 11: ...time no detected movement above the necessary movement level After every detected movement about the necessary level the counter will be reset and the movement detection starts from zero minutes agai...

Страница 12: ...Warnings In opposite to the alarm warnings will be only shown visually and not acoustically The number three 3 warnings can be configured also as acoustically or nurse call system Acoustic if enabled...

Страница 13: ...adjusted 9 3 2 Logic nurse call by this button the nurse call logic hardware relais can be permanently turned one or off If the nurse call logic is deactivated the relais will be activated in the case...

Страница 14: ...d Time 9 5 Date setup Please see chapter 10 Set up Date and Time 9 6 Touch calibration by using this function the touch screen will be calibrated The operator needs to the crosses in the display withi...

Страница 15: ...Page 15 of 23 USER MANUAL ADS intelligent Softwareversion 0 95 Inside this submenu the operator can select one language...

Страница 16: ...mode 9 8 1 Bedding mode In this mode the mattress will be inflated with full pressure for maximum 20 minutes After 20 minutes the system turns into the previous mode automatically again 9 8 2 head en...

Страница 17: ...s shown in the display and in the schematic mattress is marked by a small red bar above the motion monitor graph 1 By touching the time graph the operator can scroll horizontally from early to late in...

Страница 18: ...e current figure which is changed is highlighted in black colour 2 By pressing this button 2 the new settings will be saved 3 By pressing this button the operator will leave this submenu without savin...

Страница 19: ...patient monitoring a without redistributing any movement of patient please see chapter Error Reference source not found for details AND b Alternating cycle time mode is enabled One cycle of alternati...

Страница 20: ...le format and using the data The data will be saved in CSV format and can be opened at the PC by Excel of Calc One file per day will be saved with the name year month day csv Example a data file with...

Страница 21: ...Alarme sind m glich Warnings Priority Pump up 1 The mattress is still into pump up procedure and it is not allowed to place the patient onto the mattress If the pressure inside each cell is sufficient...

Страница 22: ...following medical device Product name ADS intelligent Meets the requirements of the European directive for medical products 93 42 EWG and 2007 47 EC The ADS clinic is consistent with risk class 1 bas...

Страница 23: ...USER MANUAL ADS intelligent Softwareversion 0 95 manufacturer of ADS intelligent AirMed PLUS GmbH Harpener Feld 34 D 44805 Bochum Germany Tel 0049 0 234 91773320 Fax 0049 0 234 91773322 Email service...

Отзывы: