If the audible alarm goes off, it can be turned off, by pressing the alarm off button
(7)
. The
alarm clock sign or triangle warning sign with exclamation mark in the display ( ) remains.
An optical alarm remains active as long as, until the cause of the alarm is resolved and the
calculated pressure value of the patient’s weigh has been reached. The acoustic alarm occurs
only once per error message.
If an optical alarm (clock symbol) occurs in case of pressure problems, please search for
a possible leakage first. The leakage may have arisen in response to one of the following
reasons:
No AD-mattress connected.
CPR-vent is open.
One or more cells are for non-pressure lie are removed and the hose connection is not
connected with the exit cover.
The hoses are not fastened with the pneumatic plug.
O-rings are defect or missing entirely up to the pneumatic plug.
The mattress has a hole or a crack.
Other leaks.
If you've checked all possible causes and you cannot find the cause, contact our service
department please.
If an optical alarm for device issues occurs, please disconnect the device from the power
supply and don’t use the device anymore. Please contact our service department.
Please note:
Alarm-function will only work when ADS clinic is conneted to the electricity supply
system. If the electricity supply system has been switched off or has failed and the
ADS clinic was temporarily switched off, you have to turn the ADS clinic on again to
make sure, that the alarming system of the ADS clinic will work fine.
Cleaning / Disinfection
For the killing of germs and bacteria, the device can de disinfected with a suitable
disinfectant. The company AirMedPLUS uses itself the product ANTISEPTICS Poly-Alcohol
skin-colorless-antiseptic.
Please note:
There should be no moisture in the device.
The AD-mattress and the cover can be disinfected by the user. For killing of germs and
bacteria, the AD-mattress can be disinfected with a cloth and a suitable disinfectant product,
egg. With a means as mentioned above. The AD-mattress included the tracing (cover) or the
tracing only, are washed in a washing machine. You should take into account the using of a
suitable detergent for killing germs and bacteria and a thermal 60 º C washing program.
E
.g. the detergent ELTRA of the company Henkel/Ecolab. Note the symbols of the AD-
mattress and cover
.
The AD-mattress and the cover should not be washed hotter than 65°. The material may
otherwise become damaged. For each re-use of the device and AD-mattress, it must be
cleaned and disinfected.