Please note:
The pressure value for each patient is at the ADS
clinic
after the test schedule for
establishing the value of the pressure relief, published by the dome of the legal
health insurance funds (GKV Union-Germany), determined. In the test report (prior
art) it is assumed that a patient, in central back position on the mattress and the
patient possesses all limbs.
If the patient is on an ADS
clinic
mattress, but now not centrally on the
mattress or as well as nursing tie in the embryonic position and only on a
higher or lower scope of the mattress is, must be the setting of the weight
of the patient, if necessary, be corrected, otherwise the patient get bed so-
res.
The cause of having bed sores can be cause by moving patients weight on a smaller
area, quasi a pressure point causes.
Display of operating hours
By pressing key
(8)
, the operating hours are displayed in whole hours. The operating hours
counter, can be turned back only by the service department of AirMedPLUS GmbH.
Turning off the device
To turn of the device, press for 2 seconds key
(2)
. Key (2) can also be pressed in case the
key security is switched on. The power-off function will take precedence over the security key.
The alarm
The ADS
clinic
is equipped with an optical and visible alarm. Early warning is monitored by the
microcontroller. The optical alarm in case of device issues, will be shown in the display as a
‚clock‘. The symbol of the optical alarm, in case of device problems, is shown by a triangle
with an exclamation mark. The acoustic alarm is an alternating beep.
'Pressure too low’
(2)
will be displayed via the LED indicator, in case there are pressure
issues. The alarm caused by a leakage or by a not yet air-filled patient’s mattress, is a safety
precaution against bed sores. The alarm, caused by a leakage, is a security against bed sores.
When using patients mattress type 1, it is necessary to insert a mattress between patient’s
mattress type 1 and the slatted/rack.
Symbol of the optical alarm at pressure issues:
This symbol will be shown on behalf of one of the following causes:
Leakage in the pneumatic system
Mattress not connected
Leakage in one of the cells or connection hose or in the device.
Symbol of the optical alarm at problems with the device:
An audible alarm is always together with a visible alarm on.