page 15 of 22
Version 1.4 – June 2015
- ADS_Clinic Operator Manual _V1_4_Mai 2015_Software 1_20_ENG.doc
Pressure relief position
In pressure relief position of particular body parts, in operating mode ‘Static’, cells can be
removed individually. To do this, you have to remove the cover of the mattress, pull out
the push buttons from the cells to be removed, left and right from the base mattress.
Tear the connection hose of the cell, to the hose system in the mattress base loose. You
now close, with the exit cover (article number 10-8007) the resulting opening at the hose
system, in order to avoid a leakage including its alarm. Pull, the temporarily cells pulled
off, from the cell regulation and connect the cover to the mattress (base system).
At a pressure relief position, in operating mode ‚Alternating pressure‘, double as large,
not filled, free areas may arise, in case the adjacent cell for non-pressure position remo-
ved cell, for the duration of the interval cycle is not filled. In this case, the patient bed
sores.
Security key
Symbol for active security key in the display:
The security key will activate automatically after ca. 5 minutes. You will recognize the
active security key by the symbol, as shown in the display. Pressing a key, in case the of
an activated security key, please follow the instructions shown in the display.
You will deactivate the security key by pressing button
(9).
If the security key in the
display is not visible, you can adjust the settings. The security key is activating itself,
after ca. 5 minutes after the last touch of the key or by pressing key
(9)
.
The security key function is to prevent unintended operation of the device. Turning off
the device is not prohibited by the security key
(2)
. Nevertheless, the user should press
key
(2)
for 2 seconds consciously, to turn off the device
Set the interval time
Setting the interval time is only possible in operating mode ‘Alternating pressure’. An
interval time from 3 till 30 minutes can be set, each time with a step of one minute. The
interval time is needed to switch cell to cell.
The desired interval time is set, by pressing key
(12)
. For a shorter interval period press
'-' for a longer interval period press '+'. The period of time interval is displayed in minu-
tes on the display.
Set the weight of the patient
Setting the weight of the patient is both possible in operating mode ‚Alternating pressure‘
and ‚Static‘.
The weight of the patient is set by key
(10)
. The steps are based on entire kilograms.
For a lower weight of the patient press '-' and for a higher weight of the patient, press
'+'.