User manual
–
HEPA
Bulle
–
ED02_09-August-2021
24 /31
3.6.1.4
After-filter removal procedure
IMPORTANT:
DANGER! BEFORE WORKING ON THE
HEPA
Bulle
UNIT, SWITCH
OFF THE DEVICE, UNPLUG THE UNIT’S POWER CABLE AND TAKE
ALL NECESSARY PRECAUTIONS TO PREVENT ANY INADVERTENT
RESTART.
-
Lay the
HEPA
Bulle
unit on its
side to clear access to the filter
panel. Make sure to lay the
machine down on a clean, flat
surface. Take precautions to
avoid scratching it.
-
Unscrew the 4 clamping knobs.
-
Remove the panel.
-
Pull the 4 HEPA filters from
their
housing
using
their
handles.
3.6.1.5
Fitting operation
Carry out the fitting operations in the opposite order to removal and reset the filter counter (see
§3.6.3 - Resetting the counter).
NOTE:
Make sure to install the filters in the correct direction: the arrow on the filter’s
label indicates the airflow direction (see §1.6.3).
3.6.2 REPLACING A FUSE
Two fuses (1.6A T) protect the electric circuit components. They are located behind the electronic
components access panel in their fuse holders (marker 7 §1.5.2).