13 - Utility menu
Factory Default
78
TGR2050 Series Instruction Manual
Factory Default
To set the instrument/ system or I/O back to default instrument settings press
UTILITY, followed by
Factory Defaults
.
The options are
Inst.
,
System
,
I/O
,
All
or
Cancel
.
Inst.
(Instrument settings) include all user set values of frequency and amplitude, all
user set operating modes including modulation type and sweep type and all the
numerical and operating settings within them.
System
(System settings) are as listed on the Utility/System menu such as screen
brightness and buzzer settings but excluding Power–on state and Date & Time.
I/O
(Input/Output settings) are those of the rear panel remote control interfaces.
All (All Factory Settings) includes all of the above.
For factory default setting details, see Appendix 1.
CAUTION
Ensure all necessary settings are saved before setting to factory default, all selected
parameters will be reset.
NOTE
The instrument will restart when set to factory default settings.
NOTE
Factory default does not affect calibration. To reset calibration to default settings,