18 - Specification
Protection
97
SMU4000 Series Instruction Manual
Protection
Over Current
Protection (OCP)
Bipolar protection limit with single setting
User settable 10µA to 3.15A
Accuracy same as current measurements ±1 digit
Or can be disabled
5 digit Resolution
Setting independent of current range
Over Voltage
Protection (OVP)
Bipolar protection limit with single setting
Accuracy same as voltage measurements ±1 digit
Or can be disabled
5 digit Resolution
Setting independent of voltage range
SMU4001
User settable 10mV to 21V
SMU4201
User settable 10mV to 210V
Over Power Protection
(OPP)
Bipolar protection limit with single setting
User settable 1µW to 26.25W
Accuracy same as power measurements ±3 digits
Or can be disabled
5 digit Resolution
Over Temperature
Protection (OTP)
Internal Temperature safety trip
Not user settable, cannot be disabled
Trip Delay (typical)
≤200us
Trip Status
Output is disabled, off condition is set to OPEN.
Measurement Result Buffer
Contents (each entry)
Index
Primary Measurement
Secondary Measurement
Date
Time (1 second resolution)
.CSV file contents
Index
Active Mode
Output state
Measurement results (V and A)
Calculated results (R and W)
System Date
System Time
Measurement Timestamp
Sequence Mode Step and Iteration
Length
Maximum of 100000 entries, FIFO style