18 - Specification
Output Power, Voltage and Current Capability
96
SMU4000 Series Instruction Manual
Source Supplementary Characteristics
Temperature coefficient (5⁰
–
18⁰C
and 28⁰
–
40⁰C)
± (0.15 x accuracy spec) / °C
Output power
25W four quadrant source or sink operation
Overrange
105% of source and measure range
Voltage Limits (compliance)
Bipolar limit with single setting
Maximum includes 105% overrange
Minimum is 10% of selected manual range
Accuracy same as voltage measurement
5 digit Resolution
SMU4001
User settable 2mV to 21V
SMU4201
User settable 2mV to 210V
Current Limits (compliance)
Bipolar limit with single setting
User settable 20nA to 3.15A
Maximum includes 105% overrange
Minimum is 10% of selected manual range
Accuracy same as current measurement
5 digit Resolution
Voltage Dropout (load modes)
Accuracy same as voltage measurements ±1 digit
5 digit Resolution (100µV fixed)
5% Hysteresis (1mV minimum)
Only available in load modes (LC, LP, LR)
Voltage range forced to auto in load modes
SMU4001
User settable up to 21V
SMU4201
User settable up to 210V
Pulse Width
User settable 200µs minimum with 10µs resolution
Wideband Noise (Voltage source)
SMU4001
<2mV (RMS, 10Hz – 20MHz)
typical into a resistive load
SMU4201
<21V: <2mV (RMS, 10Hz – 20MHz)
21V to 210V: <3.5mV (RMS, 10Hz –
20MHz)
typical into a resistive load
Source Settling Time Delay
User settable 100µs minimum with 10µs resolution
Output Power, Voltage and Current Capability
PowerFlex
PowerFlex technology enables the instrument to achieve full instrument output power across
the majority of the voltage range. 3A current limit, limits power at absolute voltages < |8.33V|.
Absolute voltages above |21V| are only achievable on the SMU4201. Output current limited to
0.5A at absolute voltages > |21V|.