AII AII-3000 A Скачать руководство пользователя страница 1

2855 Metropolitan Place, Pomona, CA 91767 USA   

   Tel: 909-392-6900, Fax: 909-392-3665, www.aii1.com, e-mail: [email protected]     Rev 6/15 

Technical Specifications 

 

 

Accuracy: 

< +2% of FS range under constant conditions 

Analysis: 

0-100% oxygen 

Application: 

Personnel safety 
Area monitoring 

O2 deficiency in confined spaces 
Checking breathing air tanks  

Checking tanks intended for scuba diving 
Confirming the O2 levels prior to welding 

Approvals: 

ISO 9001:2000, MDD 93/42/Annex II, ISO 13485:2003 

Area Classification: 

General purpose 

Alarms:   

A models: None; M models: User adjustable HI 1-100% 
and LO 0-99% alarms; 120 second alarm silence for cali-

bration purposes; HI alarm defeat for 100% O

 analysis 

Calibration: 

Air or certified 100% O

2

 every 8 hours 

Compensation: 

Temperature compensated 

Connections: 

A/M models: 1x16 mm thread; HC models: Tubing 1/4”

 

Controls: 

Soft touch keypad for ON/OFF and menu function 

Dimensions: 

3.6 x 5.9 x 1.6”; weight 10 oz. (280 grams)

 

Display: 

3-1/2 digit backlit LCD 2.75 x 1.375; resolution 0.1% O

2

 

Flow Sensitivity: 

None between 0.2 to 10 liters per minute 

Humidity: 

Non-condensing 0-95% RH 

LED Indicators: 

A models: None; M models: upon activation of alarms 

Linearity: 

+ 1% under constant conditions 

Pressure: 

Inlet - A/M models ambient, HC models - regulate; Vent 
all models - atmospheric 

Power: 

2 AA Alkaline batteries; 1,200 hours continuous use 

Response Time: 

90% of final FS reading in 9 seconds 

Sensitivity: 

< 0.5% of FS range 

Sensor: 

A/M models: AII-11-60; HC models: AII-11-60-HC  

Sensor Life: 

60 months in air at 25ºC and 1 atmosphere 

Signal Output: 

None 

Storage Temp.: 

-20º to 60ºC (-4ºF to 140ºF) on intermittent basis  

Temp. Range: 

5º to 45ºC  (41ºF to 113ºF) 

Warm-up Time: 

None 

Warranty: 

12 months analyzer; 12 months sensor  

AII-3000 A 

AII-3000 M 

ISO 9001:2008 Certified

 

INTERTEK Certificate No. 485 

  

Accessories - see back page

 

AII-3000 MHC 

AII-3000 AHC 

姚经理19961810948(微信同号)

QQ:2316517588

邮箱:[email protected]

Содержание AII-3000 A

Страница 1: ...FF and menu function Dimensions 3 6 x 5 9 x 1 6 weight 10 oz 280 grams Display 3 1 2 digit backlit LCD 2 75 x 1 375 resolution 0 1 O2 Flow Sensitivity None between 0 2 to 10 liters per minute Humidity Non condensing 0 95 RH LED Indicators A models None M models upon activation of alarms Linearity 1 under constant conditions Pressure Inlet A M models ambient HC models regulate Vent all models atmos...

Страница 2: ...HC AII 11 60 Oxygen Sensor AII 11 60 HC Oxygen Sensor BATT 1008 Battery 2x 1 5V AA BATT 1008 Battery 2x 1 5V AA P 1087 Instructions for Use P 1087 Instructions for Use A 1162 PCB Assy Main A 1162 PCB Assembly Main CABL 1006 Coil Cable TUBE 1019 7 32 OD Tubing 3 CC 1072 Carrying Case O2 姚 经 理 1 9 9 6 1 8 1 0 9 4 8 微 信 同 号 Q Q 2 3 1 6 5 1 7 5 8 8 邮 箱 2 3 1 6 5 1 7 5 8 8 q q c o m ...

Страница 3: ...e mail sales medical aii1 com web www aii1 com This manual may not be reproduced in whole or in part without the prior written consent of Analytical Industries Inc ISO 9001 2008 Certificate 485A AII 3000 AHC AII 3000 MHC AII 3000 M AII 3000 A 姚 经 理 1 9 9 6 1 8 1 0 9 4 8 微 信 同 号 Q Q 2 3 1 6 5 1 7 5 8 8 邮 箱 2 3 1 6 5 1 7 5 8 8 q q c o m ...

Страница 4: ...tanks intended for scuba diving for personnel safety area monitoring O2 deficiency in confined spaces checking breathing air tanks and confirming the O2 levels prior to welding Users must read the following statements as they are essential to re ducing the risk of use error due to ergonomic features of the device or the environment in which the device is intended to be used The devices have been d...

Страница 5: ...enting the partial pressure of oxygen in the gas stream being analyzed The resulting oxygen reading is displayed by a large easy to read backlit liquid crystal display LCD that has a resolution of 0 1 oxygen The microprocessor is controlled from a keypad and provides features like system diagnostics warning indicators controls and an alarm capability for continu ous monitoring that enhance both sa...

Страница 6: ...cluding the device is faulty If in doubt contact the manufacturer for assistance 4 NEVER operate the device in any manner described below doing so may compromise the clinical condition or the safety of patients users or other persons If the reading is unstable or a malfunction is suspected After the battery symbol appears in the LCD display Near equipment capable of emitting high levels of electro...

Страница 7: ...ot to damage the case when removing the battery compartment cover 4 2 1 Procedure 1 Remove the device and the 2 AA 1 5V Alkaline batteries from the foam shipping container 2 Turn the device over so the shortest raised line on the battery compartment cover is pointing away from you 3 Lift the tripod wire stand up and away from the case 4 Grasp the case with both hands use your thumbs press down fir...

Страница 8: ...0 AHC MHC When the HC hose connection version is ordered the device is shipped with the sensor installed 8 4 4 Controls 4 4 1 AII 3000 A AHC Oxygen Analyzers These analyzers employ a micro processor that is controlled by five 5 pushbuttons located on the keypad attached to front cover 1 ON OFF provides power to the electronics 2 ESCAPE aborts a previous selected option 3 ENTER selects a menu optio...

Страница 9: ... A AHC Oxygen Analyzers The LCD alternately displays the following until the problem is corrected Corrective action 1 Expose the sensor to air or a gas containing approximately 20 9 oxygen 2 Connect or replace the cable connecting the sensor to the analyzer 3 Connect or replace the oxygen sensor 4 5 2 2 AII 3000 M MHC Oxygen Analyzers Performs the same routine and requires the same corrective acti...

Страница 10: ...he LO alarm and proceed to the HI alarm or press the UP DOWN arrow keys to change the alarm set point 6 Press ENTER to save LO alarm set point and move to select the HI alarm 7 Press ENTER to skip the HI alarm and return to SAMPLING mode or press the UP DOWN arrow keys to change the alarm set point 8 Press ENTER to save HI alarm set point and return to SAMPLING mode 9 If no key is pressed within 5...

Страница 11: ...ixty 60 seconds to allow the sensor to stabilize before the microprocessor takes the final reading 14 7 If the calibration is successful the display below left appears for three 3 seconds before defaulting to the display below right 8 The display above right requires a decision by the user refer to warnings at the beginning of section 4 7 to press ENTER and skip the 100 O2 calibration and return t...

Страница 12: ...he sensor itself requires no maintenance and is simply replaced at the end of its useful life like a battery Inasmuch as the sensor is a transducer in its own right its ex pected life is not affected by whether the analyzer is ON or OFF The relationship between the sensor s signal and changes with the oxygen concentration is both proportional and linear thus allowing single point calibra tion Othe...

Страница 13: ...he temperature and pressure altitude at which the ana lyzer will be operated 18 Effect of Humidity The analyzer is not affected by non condensing relative humidity RH Howev er the use of a humidifier to introduce water vapor and increase the moisture level of the gas mixture does affect the oxygen concentration and the resultant reading displayed by the analyzer The addition of water vapor increas...

Страница 14: ...g displayed by the LCD If placing the entire sensor inside the controlled atmosphere review Section 5 2 Application Consideration Effect of Temperature 5 4 3 AII 3000 AHC and MHC Integral Oxygen Sensor AII 3000 AHC and MHC with their integral oxygen sensor requires connecting the tubing supplied section 4 2 1 above with the device to a hose barb attached to a pressure regulator controlling a sourc...

Страница 15: ... simply find and register the male plug at the end of the coiled cable and insert it into the mating female jack at the rear of the sensor until it mates or snaps into place 3 Calibrate the device after replacing the oxygen sensor 6 3 2 Procedure AII 3000 AHC and AII 3000 MHC Integral Sensor 1 Tools required small bladed screwdriver 2 Place the device face down on a flat surface 3 Remove the two 2...

Страница 16: ...ction Reading displayed by LCD drifts during calibration Wait 5 minutes and repeat calibration with sensor placed on flat surface not in your hand Check integrity of gas delivery system Check sensor s front o ring seal Verify calibration gas in not humidified Remove moisture covering sensor Replace sensor repeat calibration Analyzer reading climbs after calibration in 100 dry oxy gen when exposed ...

Страница 17: ...ration Air or 100 oxygen after 8 hours of continuous use Compensation Temperature compensated Connections A M models 1x16mm thread HC models Tubing 1 4 Controls Soft touch keypad for ON OFF and menu function Dimensions 3 6 x 5 9 x 1 6 weight 10 oz 280 grams Display 3 1 2 digit backlit LCD 2 5 x 1 5 resolution 0 1 O2 Flow Sensitivity None between 0 2 to 10 liters per minute Humidity Non condensing ...

Страница 18: ... or in use The purpose of the Recommended Storage Life period is to ensure the user derives the Expected Life of 60 months 915 420 oxygen hours and does not lose the benefit of the warranty Warranty The 12 month 183 084 oxygen hours warranty period begins with ship ment from the factory and is limited to the first claim submitted is based on 8 1 Spare Parts Accessories AII 3000 A AII 3000 M AII 30...

Страница 19: ...cidental or consequential losses or damages damages resulting from alterations misuse abuse lack of proper maintenance unau thorized repair or modification of the analyzer Service Contact us between 8 00am and 5 00pm PST Monday thru Thursday or before 12 00pm on Friday Trained technicians will assist you in diagnosing the prob lem and determining the appropriate course of action 31 10 Material Saf...

Отзывы: