REC
” flashes on the meter display, above and to the left
of the temperature reading. The interval of the flashing
corresponds to the time interval set in the previous step
(6.3).
- The computer monitor then displays a green screen with
the values currently measured and a summary of the
settings selected for the recording (on the right). The
threshold values not respected are displayed in black on
the green screen (e.g.: if the threshold set for the CO2 is
1200 ppm and the measurement recorded is higher, the
threshold of 1200 ppm is highlighted in black).
Example recording:
10 readings every minute in room 1
on PC1 called Measurements (so the total recording time
is 10 minutes). Minimum temperature of 24°C and
maximum temperature of 30°C. Minimum humidity 42%
and maximum 50%. A CO2 level of 1200 ppm.
Set Param
→
Address
: “1” →
Serial Number
: “PC1” →
Test name
: “Measurements” →
Total Records
: “10” →
Interval
: “1 min” →
Temp. Low
: “24°C” →
Temp High
“30°C” →
Humidity Low
: “42%” →
Humidity High
: “50%”
→
CO2
: “1200 ppm”.
Important:
The “address” field must be numerical.
Make sure the alarm is enabled. The minimum and
maximum values for temperature and humidity are used as
reference thresholds for the chart. As indicated in step 2 of
the instructions, the alarm is only available when the set
CO2 value is exceeded.