Output modules
When configuring a measured value output it is also possible, in the same
menu, by means of the functions
Analog start
and
Analog end
, to have the
measuring range actually used for the measuring point concerned spread over
the full 10 V or 20 mA; (see 12.4.4)
To program the analog output start
:
6 Analog start :
0.0°C
To program the analog output end
(see 10.5):
6 Analog end : 100.0°C
For 20 mA analog outputs only
To choose between 0 - 20 mA and 4 - 20 mA output :
Curr. output t 4-20 mA
Programmed analog value output
(see Manual 6.10.7)
In variant 8 "Driven externally"
8: Driven externally
the analog output value can be programmed (see 10.5):
Analog value : 5.000 mA
12.7 Power supply menu
The power supply for the measuring instru-
ment is normally derived from mains adapter
ZB 1212-NA9 (12V/2.5A). There is also the op-
tion of using module ES 5690-AP with 8 AA
NiMH rechargeable batteries. The power sup-
ply menu displays the current battery voltage
to help you estimate the battery's remaining
operating time. At 10.4 V the battery symbol in
the status bar starts to flash and at 8.8 V the
device switches off automatically. The current
charge status cannot be displayed more ex-
actly than this because of the different types of
load.
Display of supply voltage / battery voltage
Battery voltage:
10.8 V
Display of actual sensor voltage
Sensor voltage :
11.6 V
12.8 Locking, calibration menu (option KL)
In the
Locking and calibration
menu you can
lock the right-of-access to certain menus and
to certain functions. Here you can also see
the serial numbers and calibration data for
the device itself and for any sensors attached.
With option KL it is possible not only to correct
the sensor at several points in the connector
itself (see 12.3.11) but also to manage the as-
sociated calibration data.
The right-of-access to this and other menus and
to certain key functions can be stipulated in de-
tail and protected by password by means of the
parameters "Menu" and "Fct".
ALMEMO
®
5690-2M
81
* Locking, calibration*
±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±
Password:
****
Locking level: Menu: 0 Fct: 0
Device:
2890-9 6.22
Serial number:
04020123
Next calibration:
01.12.05
Signal for calibration:
Ø
Sensor:
Channel: 00
Type:
FHA646-6
Serial number:
04020123
Next calibration::
01.02.06
Calibr. interval:
12 Month
PRINT ESC
* POWER SUPPLY *
¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯¯
Battery voltage
10.8 V
Sensor voltage
11.6 V
ESC