Selecting the measuring range
Sensor
Connector /
cable / sensor
Measuring range Units Display
Sensor voltage
any
0.00...20.00
V
Batt
Frequency
ZA 9909-AK
0... 25000
Hz
Freq
Pulses
ZA 9909-AK
0... 65000
Puls
Digital input
ZA 9000-EK2
0.0... 100.0
%
Inp
Digital interface
ZA 9919-AKxx
-65000... +65000
DIGI
Measured value (Mb1)
any
f(
Mb1
)
Mess
Differential (Mb1-Mb2)
any
f(
Mb1
)
Diff
Maximum value (Mb1)
any
f(
Mb1
)
Max
Minimum value (Mb1)
any
f(
Mb1
)
Min
Average value over time (Mb1)
any
f(
Mb1
)
M(t)
Number of averaged values (Mb1)
any
n(t)
Average val. over meas. points (Mb2, Mb1) any
f(
Mb1
)
M(n)
Total from measuring points (Mb2, Mb1)
any
f(
Mb1
)
S(n)
Total number of pulses (Mb1)
ZA 9909-AK
s.Man.6.7.1
0..65000
S(t)
Pulses / print cycle (Mb1)
ZA 9909-AK
s.Man.6.7.1
0..65000
S(P)
Alarm value (Mb1)
any
0/100
%
Alrm
Volume flow m
3
/h
Mb
1
· Q
any
m
3
/h
Flow
Timer 1
any
0...65000
s
Time
Timer 2 (exponent -1)
any
0.0...6500.0
s
Time
Mbx = reference channels
With the function channels it is possible to represent function parameters from
measured value processing or from calculated results obtained by linking cer-
tain measured values on measuring channels (see Manual 6.3.4). Reference to
the actual measuring channels is provided by one or two reference channels.
For all function channels the default reference channels Mb1 and Mb2 are
available on the appropriate plug; these do not need programming.
Function
Function channel
Ref. channel 1
Ref. channel 2
Function parameters (Mb1)
on channel 2, 3, or 4
Mb1 = channel 1
Differential (Mb1-Mb2)
on channel 2, 3, 4 (Mb1)
Mb1 = channel 1
Mb2=M0.0
Average value over Mb2, Mb1
on channel 2, 3, 4 (Mb1)
Mb1 = channel 1
Mb2=M0.0
Total over Mb2, Mb1
on channel 2, 3, 4 (Mb1)
Mb1 = channel 1
Mb2=M0.0
Arrangement of channels on the plugs:
Once the quantity has been programmed the default reference channels can
be used (see above). Settings for the reference channels are described in
15.12.6.
15.10 Sensor configuration
D6 and D7 sensors can be assigned new quantities, measuring ranges, and in-
dividual parameters completely unknown to the measuring instrument. Each
such sensor must therefore provide its own sensor menu for defining the sen-
sor configuration with all its special settings (i.e. measuring ranges, compensa-
ALMEMO
®
204
49
Содержание ALMEMO 204