7.2 Audit Trail Parameters
Two separate incrementing, non-resetable audit trail parameters are used by the Model
SBI-521 to indicate changes to various parameters or calibration.
7.2.1 View Configuration Counter
1.
Press and hold the
[SETUP]
key for 3 seconds.
CONFIG
is displayed
2.
Press the
[TARE]
key.
2a.
The display will briefly show the number of times the configuration has been
changed and saved up to 999.
To Exit:
1.
Press the
[ON/OFF]
key.
Exit?
is displayed.
2.
Press the
[TARE]
key to return to weigh mode.
7.2.2 View Calibration Counter
1.
Press and hold the
[SETUP]
key for 3 seconds.
CONFIG
is displayed.
2.
Press the
[FUNC]
key twice.
Cal
is displayed.
3.
Press the
[TARE]
key.
3a.
The display will briefly show the number of times the configuration has been
changed and saved up to 999.
To Exit:
1.
Press the
[ON/OFF]
key.
Exit?
is displayed.
2.
Press the
[TARE]
key to return to the weigh mode.