Step 5: Verify Operation of the U5309A Module
Step 5: Verify Operation of the U5309A Module
The intention of this step is to verify correct operation of the newly installed module. Run Agilent Connection
Expert by right-clicking the task bar icon
, and select
Agilent Connection Expert
.
It will display the modules that are installed. Review the configuration data and then launch the SFP. This will
provide control of the module for self test and other operational verification procedures.
Agilent U5309A Instrument Properties
The instrument properties may
be viewed by clicking on the
desired instrument and then on
the
Instrument Properties
tab
in the right pane.
If the module
does not appear in the Agilent
Connection Expert, first try the
'Refresh All' button. If that
does not work, restart your PC
or embedded controller and
start Agilent Connection Expert
again.
Agilent U5309A Installed Software
A verification utility may be
launched by selecting the
Installed Software
tab, selecting
the 'SFP - Verify Operation' item
and then clicking on
Start SFP
.
Agilent U5309A Startup Guide
17