![Agilent Technologies U2701A Скачать руководство пользователя страница 88](http://html1.mh-extra.com/html/agilent-technologies/u2701a/u2701a_user-manual_2868720088.webp)
2
Scope Features and Functions
72
U2701A/U2702A Modular Oscilloscopes User’s Guide
Display Options
The figure below shows the
Display Options
menu on the soft
front panel.
The three types of display options offered are:
• Vector
connects the sample points by using digital
interpolation.
• Sin x/x interpolation
is used to expand the horizontal signal
resolution when the horizontal scale is set to 100 ns or faster.
To perform interpolation using sin(x)/x filter to maintain the
linearity of the waveform, enable the
Sin x/x interpolation
option. Once enabled, a smoother waveform will be
displayed.
Figure 2-17
The waveform before interpolation
U2700A_Modular_Scope.book Page 72 Tuesday, May 6, 2008 9:32 AM