ROUTe:CHANnel:RANGe
Syntax
ROUTe:CHANnel:RANGe <value>, (@<ch_list>)
This command is used to set the voltage range of the specified analog
input channels.
ROUTe:CHANnel:RANGe? (@<ch_list>)
The query command returns a float value that represents the range in
volts of the specified analog inputs channels.
Multiple responses are
separated by commas.
Supported Model
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Parameters
Item
Type
Range of Values
Default Value
<value>
Float
U2331A: Unipolar mode: {10|5|4|2.5|2.0|1.0|0.5|0.4|0.1}
Bipolar mode: {10|5|2.5|2.0|1.25|1.0|0.5|0.25|0.2|0.05}
U2351A/U2352A/U2353A/U2354A/U2355A/ U2356A:
Unipolar/Bipolar mode: {10,5,2.5,1.25}
10
<ch_list>
Numeric
Single-ended mode:
U2351A/U2352A/U2353A/U2354A: 101-116
U2355A/U2356A/U2331A: 101-164
Differential mode:
U2351A/U2352A/U2353A/U2354A: 101-108
U2355A/U2356A/U2331A: 101-132
N/A
Содержание U2300A Series
Страница 1: ...Agilent Technologies Agilent U2300A Series Multifunction USB Data Acquisition Programming Guide ...
Страница 8: ...VIII U2300A Series USB DAQ Programming Guide Contents ...
Страница 18: ...1 Introduction to Programming 10 U2300A Series USB DAQ Programming Guide ...
Страница 40: ...5 CONFigure Subsystem 32 U2300A Series USB DAQ Programming Guide ...
Страница 60: ...6 IEEE 488 Common Command 52 U2300A Series USB DAQ Programming Guide ...
Страница 70: ...7 MEASure Subsystem 62 U2300A Series USB DAQ Programming Guide ...
Страница 84: ...8 OUTPut Subsystem 76 U2300A Series USB DAQ Programming Guide ...
Страница 92: ...9 Root Commands 84 U2300A Series USB DAQ Programming Guide ...
Страница 104: ...10 ROUTe Subsystem 96 U2300A Series USB DAQ Programming Guide See Also ROUTe CHANnel RSouRCe ...
Страница 130: ...11 SENSe Subsystem 122 U2300A Series USB DAQ ...
Страница 162: ...15 WAVeform Subsystem 154 U2300A Series USB DAQ Programming Guide ...
Страница 168: ...16 Error Messages 160 U2300A Series USB DAQ Programming Guide ...