5
CONFigure Subsystem
28
U2300A Series USB DAQ Programming Guide
CONFigure:TIMEbase:SOURce
Syntax
CONFigure:TIMEbase:SOURce <mode>
This command is used to set the timebase/clock source for both analog
inputs and outputs.
CONFigure:TIMEbase:SOURce?
The query command returns a string value, which represents the
timebase/clock source for both analog input and output channels.
Supported Model
U2351A, U2352A, U2353A, U2354A, U2355A, U2356A, U2331A
Parameters
Query Returned Format
<mode> ::= {INTernal|EXTernal|CCG}
Example
CONFigure:TIMEbase:SOURce EXTernal
CONFigure:TIMEbase:SOURce?
//Expected returned string is EXTernal
Item
Type
Range of Values
Default Value
<mode>
String
INTernal: Internal clock as the base time
EXTernal: External clock as the base time
CCG: Clock based on U2781A modular
instrument chassis
INTernal
Содержание U2300A Series
Страница 1: ...Agilent Technologies Agilent U2300A Series Multifunction USB Data Acquisition Programming Guide ...
Страница 8: ...VIII U2300A Series USB DAQ Programming Guide Contents ...
Страница 18: ...1 Introduction to Programming 10 U2300A Series USB DAQ Programming Guide ...
Страница 40: ...5 CONFigure Subsystem 32 U2300A Series USB DAQ Programming Guide ...
Страница 60: ...6 IEEE 488 Common Command 52 U2300A Series USB DAQ Programming Guide ...
Страница 70: ...7 MEASure Subsystem 62 U2300A Series USB DAQ Programming Guide ...
Страница 84: ...8 OUTPut Subsystem 76 U2300A Series USB DAQ Programming Guide ...
Страница 92: ...9 Root Commands 84 U2300A Series USB DAQ Programming Guide ...
Страница 104: ...10 ROUTe Subsystem 96 U2300A Series USB DAQ Programming Guide See Also ROUTe CHANnel RSouRCe ...
Страница 130: ...11 SENSe Subsystem 122 U2300A Series USB DAQ ...
Страница 162: ...15 WAVeform Subsystem 154 U2300A Series USB DAQ Programming Guide ...
Страница 168: ...16 Error Messages 160 U2300A Series USB DAQ Programming Guide ...