Agilent U1840A Medalist sj5000 Hardware Manual
6
Auto Width Adjust Conveyor Setup
Clamp Home Sensor Adjustment
The Clamp Home sensor has to be adjusted its fit and flatness
between the board and top finger clamp. Follow the steps below
to adjust the sensor.
The tools required are:
•
Clamping tolerance test board
•
50
μ
m filler gauge
•
Metric Allen keys
The board clamping plate is 510 mm in length. It is important to
ensure the flatness of the plate along its length when a
production board is clamped.
1
Activate Maintenance mode and manually insert the test
board onto the clamping location.
2
Click the clamps
Up
button in the Maintenance window to
clamp the board.
3
Insert the 50
μ
m filler gauge into every clamping tooth that is
in contact with the board. To pass the test, the filler gauge
should bend when you try to insert it between each clamping
tooth and the board.
4
If the clamping mechanism fails the filler gauge test, adjust
the Clamp Home sensor to enable the clamp plate to move
further up and provide a firmer clamp. To do so, loosen the
sensor screws, push up the sensor slightly and tighten the
screws.
Содержание U1840A Medalist sj5000
Страница 15: ...Safety and Regulatory Information 1 Agilent U1840A Medalist sj5000 Hardware Manual 1 9...
Страница 50: ...3 12 Agilent U1840A Medalist sj5000 Hardware Manual 3 Installation...
Страница 100: ...7 18 Agilent U1840A Medalist sj5000 Hardware Manual 7 Camera and Lighting Calibration...
Страница 134: ...10 18 Agilent U1840A Medalist sj5000 Hardware Manual 10 Setting Up Barcode Readers...
Страница 176: ...11 42 Agilent U1840A Medalist sj5000 Hardware Manual 11 Troubleshooting...
Страница 181: ...Maintenance 12 Agilent U1840A Medalist sj5000 Hardware Manual 12 5 Figure 12 2 Sensors on the conveyor...
Страница 190: ...A 4 Agilent U1840A Medalist sj5000 Hardware Manual A System Specifications...
Страница 192: ...B 2 Agilent U1840A Medalist sj5000 Hardware Manual B Reference Documents...