Agilent U1251A/U1252A User’s and Service Guide
2
Making
Measurements
Measuring Temperature
The bead type thermocouple probe is suitable for making
temperature measurements between –20 °C to 200 °C in
Teflon compatible environments. Above this temperature,
probes may emit toxic gas. Do not immerse this thermocou-
ple probe in liquids. For best results, use a thermocouple
probe designed for each application — an immersion probe
for liquid or gel, an air probe for air measurements. Observe
the following measuring techniques:
•
Clean the measurement surface and make sure the probe
is securely touching the surface. Remember to disable the
applied power.
•
When measuring above ambient temperature, move the
thermocouple along the surface until you get the highest
temperature reading.
•
When measuring below ambient temperature, move the
thermocouple along the surface until you get the lowest
temperature reading.
•
Place the meter in the operating environment for at least
1 hour as using non- compensation transfer adaptor with
miniature thermal probe.
•
For quick measurement, use the 0 °C compensation to see
the temperature variation of the thermocouple sensor. The
0 °C compensation assists you in measuring relative
temperature immediately.
1
Turn the rotary switch to
TEMP
position
.
2
Push
to select temperature measurement.
C A U T I O N
Do not bend the thermocouple leads at sharp angles. Repeated bending
over a period of time can break leads.
Содержание U1251A
Страница 1: ...Agilent Technologies Agilent U1251A and U1252A Handheld Digital Multimeter User s and Service Guide...
Страница 34: ...34 Agilent U1251A U1252A User s and Service Guide 1 Getting Started Tutorial...
Страница 37: ...Making Measurements 2 Agilent U1251A U1252A User s and Service Guide 37 Figure 3 Measuring AC voltage MAX...
Страница 43: ...Making Measurements 2 Agilent U1251A U1252A User s and Service Guide 43 Figure 8 Measuring frequency Push MAX...
Страница 47: ...Making Measurements 2 Agilent U1251A U1252A User s and Service Guide 47 Figure 11 Conductance measurement MAX...
Страница 74: ...74 Agilent U1251A U1252A User s and Service Guide 3 Features and Functions Figure 26 Log review mode operation...
Страница 112: ...112 Agilent U1251A U1252A User s and Service Guide 5 Maintenance Figure 52 Battery charging procedures...
Страница 114: ...114 Agilent U1251A U1252A User s and Service Guide 5 Maintenance Figure 53 Fuse replacement Fuse 1 Fuse 2...
Страница 116: ...116 Agilent U1251A U1252A User s and Service Guide 5 Maintenance...
Страница 120: ...120 Agilent U1251A U1252A User s and Service Guide 6 Accessories...
Страница 156: ...156 Agilent U1251A U1252A User s and Service Guide 8 Specifications...