Agilent SPS 4 Autosampler User’s Guide
3
Installation
Fitting the Sample Probe
Figure 13
Installing the sample probe
To fit the sample probe:
1
Turn the autosampler off.
2
Manually rotate the probe arm so it faces the front of the autosampler
for easy access.
3
Undo the mount nut from the probe mount block on the sample probe
and slide the nut off the sample probe.
4
Place the mount nut from the sample probe on top of the accessories
plate with the recessed edge facing up and the holes approximately
aligned.
5
Move the Z- axis slide down stopping just above mount nut sitting on
the accessories plate.
WA R N I N G
Always switch off the power to the autosampler before installing or removing the
sample probe.
Z-Axis Slide
Eyelet for tube routing
Tube routing this side
Tube restraint
Probe mount block
Mount nut
Sample probe
Accessories plate