SCPI Command Summary
68
Series N6700 User’s Guide
SCPI Command Summary
Subsystem Commands
NOTE
Some [optional] commands have been included for clarity.
SCPI Command
Description
ABORt
:ACQuire (@chanlist)
Resets the measurement trigger system to the Idle state
:TRANsient (@chanlist)
Resets the transient trigger system to the Idle state
CALibrate
:CURRent
[:LEVel] <NRf>, (@channel)
Calibrates the output current programming
:MEASure <NRf>, (@channel)
Calibrates the current measurement
:PEAK (@channel)
Calibrates the peak current limit (Note 1)
:DATA <NRf>
Enters the calibration value
:DATE <SPD>, (@channel)
Sets the calibration date
:DPRog (@channel)
Calibrates the current downprogrammer
:LEVel P1 | P2
Advances to the next calibration step
:PASSword <NRf>
Sets the numeric calibration password
:SAVE
Saves the new cal constants in non-volatile memory
:STATE <Bool> [,<NRf>]
Enables/disables calibration mode
:VOLTage
[:LEVel] <NRf>, (@channel)
Calibrates the output voltage programming
:CMRR (@channel)
Calibrates the voltage common mode rejection ratio (Note 1)
:MEASure <NRf>, (@channel)
Calibrates the voltage measurement
FETCh (Note 2) | MEASure
[:SCALar]
:CURRent [:DC]? (@chanlist)
Returns the average output current
:VOLTage [:DC]? (@chanlist)
Returns the average output voltage
:ARRay
:CURRent [:DC]? (@chanlist)
Returns the instantaneous output current (Note 2)
:VOLTage [:DC]? (@chanlist)
Returns the instantaneous output voltage (Note 2)
INITiate
[:IMMediate]
:ACQuire (@chanlist)
Enables the measurement system to receive triggers (Note 2)
:TRANsient (@chanlist)
Enables the output transient system to receive triggers
:CONTinuous
:TRANsient <Bool>, (@chanlist)
Enables/disables continuous transient triggers
OUTPut
[:STATe] <Bool>, [NORelay], (@chanlist)
Enables/disables the specified output channel(s)
:DELay
:FALL <NRf+>, (@chanlist)
Sets the output turn-off sequence delay
:RISE <NRf+>, (@chanlist)
Sets the output turn-on sequence delay
:INHibit
:MODE LATChing | LIVE | OFF
Sets the remote inhibit input
:PON
:STATe RST | RCL0
Programs the Power-On State