98
PL-GPC 220 User Manual
7
Maintenance
Autosampler
Lubricate Autosampler Syringe (not applicable to serial controlled
A/S systems)
1
Pull out the A/S carousel drawer to its maximum position to gain access to
the syringe assembly beneath it.
2
Reach in over the top of the extended carousel and with both hands lift the
syringe from its mounting. Disconnect it from the Luer lock needle
assembly.
3
If the syringe is contaminated with solvent, wash it out with acetone and
dry.
4
Place a couple of drops of silicone oil into the barrel of the syringe and
replace the plunger.
5
Work the plunger up and down the barrel a few times to thoroughly
disperse the oil.
6
Reconnect the syringe to the tubing and relocate in the mounting.
7
Return the A/S drawer to its normal operating position.
Parts required
Description
Acetone
Silicone oil