94
Installation and Verification Manual
3
Performance Tests
13
Ensure the OmniBER OTN
Laser
is
ON
.
14
Select
Eye/Mask Mode
,
Mask Test.
Select
Filter
and ensure
it is set to
OC-192 (9.95Gb/s), ON
15
Change the mask to OTU-2.
16
Select oscilloscope
Autoscale
and allow the oscilloscope to
find signal and trigger inputs
17
Start
Mask Test
and check the Optical Output meets the
OTU-2 mask on
18
Set the oscilloscope to
Eye Measure
mode and setup to
measure the
Average Power
(in dBm) and
Extinction Ratio
.
19
Check the Average Optical Power is between
–1 to +1 dBm
for
version 1 optics OR
–5 to –1
dBm
for version 2. Check the
Extinction Ratio is greater than
8.2 dB
and record the
measured value.
20
Tests Complete. Set the Laser to OFF before disconnecting
setup.
N O T E
Ensure Vertical Calibration and Extinction Ratio Calibration have been
performed for the oscilloscope channel and vertical sensitivity range to
ensure measurement accuracy. Refer to oscilloscope information/Help if
necessary.