124
Installation and Verification Manual
3
Performance Tests
3
Connect the
8 Mb/s OUT
port to the Oscilloscope CHAN 1, using
the E2622A Termination Adapter Probe.
.
4
Load the
8Mb_ITU.msk
test mask and check the Positive and
Negative pulses are within the mask (see examples).
Adapter
54810A Oscilloscope
OmniBER OTN
CH1
8 Mb/s Output
BNC to BNC cable
E2622A
75/50
Ω
8 Mb/s O/P
N O T E
To test Negative pulse set the Scope
Mask Test Set-up
panel to
Invert
Mask
panther3_iv.book Page 124 Wednesday, January 15, 2003 12:03 PM