
116
Installation and Verification Manual
3
Performance Tests
3
Connect the OmniBER OTN’s
DS1 OUT
port to the
Oscilloscope using the E2621A Termination Adapter Probe,
set probe switch to
100
Ω.
4
Load the
DS1_ANSI.msk
test mask and run the mask test. Check the
Positive and Negative pulses are within the mask (see examples).
Bal/Unbal
Adapter
54810A Oscilloscope
OmniBER OTN
CH1
E2621A
DS1 Balanced O/P
Set
to
100
Ω
Bantam to Bantam Cable
DS1 O/P
N O T E
To test Negative pulse set the Scope
Mask Test Set-up
panel to
Invert
Mask.
panther3_iv.book Page 116 Wednesday, January 15, 2003 12:03 PM
Содержание OmniBER J7230A
Страница 2: ...sdh_Lynx2 book Page 148 Wednesday April 17 2002 12 49 PM...
Страница 10: ...8 Installation and Verification Manual panther3_iv book Page 8 Wednesday January 15 2003 12 03 PM...
Страница 172: ...170 Installation and Verification Manual panther3_iv book Page 170 Wednesday January 15 2003 12 03 PM...
Страница 173: ...sdh_Lynx2 book Page 148 Wednesday April 17 2002 12 49 PM...