NovoCyte Quanteon Technical Specifications
NovoCyte Quanteon Flow Cytometer Operator's Guide
143
Appendix B:
NovoCyte Quanteon Technical
Specifications
This chapter provides technical specifications of NovoCyte Quanteon flow
cytometer and associated standard accessories.
NovoCyte Quanteon Technical Specifications
Table 15 NovoCyte Quanteon Technical Specifications
1
Optics
Laser
Solid state laser with on-board thermal-electric
cooling and guaranteed thermal stability and
beam quality
Laser Power
100 mW
Laser Beam Profile
10 x 60 µm elliptical beam
Optical Alignment Procedure
Fixed, no operator alignment required
Fluorescence Detection
Silicon Photomultiplier (SiPM) with high photon
detection efficiency; Individual photodetector for
each channel
FSC/SSC Sensitivity
FSC: 0.4 µm; SSC: 0.1 µm
Fluorescence Resolution
<3% CV for CEN
Optical Filters
User exchangeable, “Smart” filter automatically
read by the system
Cell Size
0.1 - 50 µm
Fluidics
Flow Cell
170 x 290 µm rectangular quartz flow cell
Sample Acquisition Rate
100,000 events/second
Sample Delivery
Positive-displacement syringe pump enabling
direct absolute count without the need for
reference counting beads.
Volumetric Absolute Count
Precision
< 5%
Содержание NovoCyte Quanteon
Страница 10: ...Contents NovoCyte Quanteon Flow Cytometer Operator s Guide 10 This page is intentionally left blank...
Страница 28: ...Prologue 28 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 70: ...Instrument Operation 70 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 106: ...Running Samples 106 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 136: ...Troubleshooting 136 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 138: ...Version History 138 NovoCyte Quanteon Flow Cytometer Operator s Guide This page is intentionally left blank...
Страница 147: ......