Chapter 2
83
Test Descriptions
Measurement Jitter
Test Procedure
NOTE
Throughout the Measurement Jitter Test Procedure the term ‘the Worksheet’ refers
to the Measurement Jitter Test Worksheet on page 86. The term ‘the Test Record’
refers to the Measurement Jitter Test Record detailed in the relevant model number
Appendix for the being tested.
Step 1.
Connect the equipment as shown in Figure 2-24.
Step 2.
Ensure the preset is set to factory settings (press the
System
key, then
More
,
Power
On/Preset
,
Power On (Preset)
and
Preset (Factory)
).
Step 3.
Press the
Preset
key on the NFA and wait for the routine to finish.
Step 4.
Load the noise source diskette (A:) or from the NFA’s internal memory (C:).
Alternatively, create an ENR file from the data supplied on the noise source label
(refer to the
Noise Figure Analyzers NFA Series User's Guide
) for more details). Do
not calibrate the source. The NFA should display
Uncorrected
in the active
window.
Step 5.
Set the NFA as follows:
1. Set the measure result to Y-Factor (press the
Result
key, then
Y-Factor
).
2. Set sweep to single (press the
Sweep
key, then
Sweep Mode: single
).
3. Set the frequency mode to fixed (press the
Frequency/Points
key, then
Freq
Mode
and
Fixed
).
4. Set the fixed frequency to 1 GHz (press
Frequency/Points
,
Fixed Freq
,
1
, GHz.
5. Set measure to Linear mode (press the
Scale
key, then
Units Linear
).
6. Select the display to meter mode (press the
Format
key, then
Format
and
Meter
).
Step 6.
Restart the NFA sweep (press the
Restart
key).
Step 7.
Repeat step 6, recording 100 displayed linear Y-Factor values within the Worksheet.
Step 8.
Calculate the linear standard deviation with the equation given below:
Содержание NFA Series
Страница 25: ...13 2 Test Descriptions This chapter descibes each test required for calibration and performance verification...
Страница 114: ...104 Chapter3 Technical Specifications General Specifications...
Страница 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Страница 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Страница 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Страница 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Страница 184: ...174 Index Index...