50
Chapter 2
Test Descriptions
Frequency Accuracy
Step 6.
Use Cells D1 to D201 as a look up reference and find the cell, which closest
matches the ½ Power Reading in cell D202.
The corresponding frequency in Column A is given as the center frequency point.
Enter the reading into the appropriate column in the Test Record. Ensure that the
measured value meets its published specification.
Performing the frequency accuracy calculations manually
Step 1.
Using the displayed P
hot
results sum the values and enter the summed values on the
Test Worksheet Table on page 45.
Example 2-2
1 GHz CW at the 4 MHz bandwidth
For the 1 GHz CW at the 4 MHz bandwidth each Phot reading shown in the Figure
2-12 on page 51 is summed with the previous cell(s) until the total power under the
curve is equated in the lower right hand column. Example sum reading 1 to reading
6 = 0.369.
Figure 2-11
Displayed Table Values
Содержание NFA Series
Страница 25: ...13 2 Test Descriptions This chapter descibes each test required for calibration and performance verification...
Страница 114: ...104 Chapter3 Technical Specifications General Specifications...
Страница 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Страница 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Страница 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Страница 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Страница 184: ...174 Index Index...