![Agilent Technologies NFA Series Скачать руководство пользователя страница 61](http://html1.mh-extra.com/html/agilent-technologies/nfa-series/nfa-series_performance-verification-and-calibration-manual_2868187061.webp)
Chapter 2
49
Test Descriptions
Frequency Accuracy
Performing the frequency accuracy calculations using the CSV file in a
spreadsheet
The following procedure uses Microsoft Excel.
Step 1.
Open Excel.
Step 2.
Open the .csv file from the A: drive. From the
File
menu select
Open
and
Look in:
(A:). Select files of
Type:
Text Files (*.prn; *.txt; *.csv).
Figure 2-10
Example Excel File
Column A contains the measured frequency. Column B contains the corresponding
Linear P
hot
value. Disregard Column C null terminator information.
Step 3.
Increase the column widths to 20 to view the data more accurately.
Step 4.
Enter the following formula into cell D1 = SUM(B$1:B1), copy this formula and
paste it from cell D2:D201.
Step 5.
Enter the following formula into cell D202 =D201/2. This value returns the ½
Power reading.
Содержание NFA Series
Страница 25: ...13 2 Test Descriptions This chapter descibes each test required for calibration and performance verification...
Страница 114: ...104 Chapter3 Technical Specifications General Specifications...
Страница 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Страница 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Страница 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Страница 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Страница 184: ...174 Index Index...