![Agilent Technologies NFA Series Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/agilent-technologies/nfa-series/nfa-series_performance-verification-and-calibration-manual_2868187051.webp)
Chapter 2
39
Test Descriptions
Frequency Accuracy
Related Adjustment
None
Required Test Equipment
NOTE
Synthesized Sweeper model 83620B cannot be used test NFA N8975A.
Table 2-7
Equipment required for Frequency Accuracy test
Equipment description
Critical specification for equipment
substitution
Recommended model
Synthesized Sweeper
(see note)
Frequency Range: 10 MHz to 26.5 GHz
Frequency Accuracy (CW): 0.02%
Power Level Range: -55dBm
83620/30/40/50B Option
001 and 008
Adapter 1
N8973A N Type (M) to 3.5mm (F)
N8974/5A 3.5mm Precision (F) to 3.5mm
(F)
1250-1744
1250-1749 or 83059B
Adapter 2
83620/30 3.5mm Precision (F) to 3.5mm (F)
83640/50 2.4mm Precision (F) to 3.5mm (F)
1250-1749 or 83059B
11901B
RF Cable
10 MHz to 26.5 GHz
3.5mm (M) to 3.5mm (M), 61 cm (24 in.)
11500E
Frequency Standard
Frequency: 10 MHz Timebase Accuracy
(Aging): < 1 x 10
-10
/day
5071A
BNC Lead (X 2)
Length 122cm (48 in.), Frequency 10 MHz
10503A
Содержание NFA Series
Страница 25: ...13 2 Test Descriptions This chapter descibes each test required for calibration and performance verification...
Страница 114: ...104 Chapter3 Technical Specifications General Specifications...
Страница 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Страница 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Страница 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Страница 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Страница 184: ...174 Index Index...