![Agilent Technologies NFA Series Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/agilent-technologies/nfa-series/nfa-series_performance-verification-and-calibration-manual_2868187043.webp)
Chapter 2
31
Test Descriptions
Input VSWR
Measuring VSWR from 10 MHz to 500 MHz
Step 1.
Connect the test equipment as shown in Figure 2-4 on page 29.
Step 2.
Set the Network Analyzer Start/Stop frequencies from 10 MHz to 500 MHz.
Press the
Start
key then
1, 0,
M/µ;
then
press the
Stop
key and
5, 0, 0
M/µ
.
NOTE
Ensure the VNA
Correction
and
Interpol
functions are both set to
On
.
Step 3.
On the NFA ensure that the preset is set to the factory settings (press the
System
key
and
More
,
Power On/Preset
,
Power On (Preset)
and
Preset (Factory)
).
Step 4.
On the NFA press the
Preset
key and wait for the routine to finish
Step 5.
Set sweep to single (press the
Sweep
key and
Sweep Mode: Single
).
Step 6.
Set the frequency mode to fixed (press the
Frequency/Points
key and
Freq Mode
,
Fixed
).
Step 7.
Set the fixed frequency to 250 MHz (press the
Frequency/Points
key and
Fixed
Freq
,
2, 5, 0
,
MHz
).
Step 8.
Restart the measurement sweep on the DUT (press the
Restart
key).
Step 9.
Set the Network Analyzer to autorange (press the
Scale Ref
key and
Auto Scale
).
Step 10.
Using the marker peak function on the network analyzer, determine the maximum
displayed VSWR over the measured frequency range (press the
Marker Fct
n key
and
Mkr Search [ON]
,
Search: Max
).
Step 11.
Record the maximum displayed signal frequency and VSWR on the Test Record.
Step 12.
Verify that the measured VSWR from 10 MHz to 500 MHz is within the published
specification.
Содержание NFA Series
Страница 25: ...13 2 Test Descriptions This chapter descibes each test required for calibration and performance verification...
Страница 114: ...104 Chapter3 Technical Specifications General Specifications...
Страница 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...