Caring for Connectors
Using Connector Gages
•
Gently rock the two surfaces together, to make sure that they have come together
flatly. The gage pointer should now line up exactly with the zero mark on the
gage. If it does not, inspect and clean the gage and gage calibration block again
and repeat this process. If the gage pointer still does not line up with the zero
mark on the gage, loosen the dial lock screw and turn the graduated dial until the
gage pointer exactly lines up with zero. Then retighten the lock screw.
Figure D-6
Using the connector gage
NOTE
Gages should be checked often, to make sure that the zero setting has not changed.
Generally, when the gage pointer on a gage that has been zeroed recently does not
line up exactly with the zero mark, the gage or calibration block needs cleaning.
Clean carefully and check the zero setting again.
Содержание NFA Series
Страница 25: ...13 2 Test Descriptions This chapter descibes each test required for calibration and performance verification...
Страница 114: ...104 Chapter3 Technical Specifications General Specifications...
Страница 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Страница 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Страница 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Страница 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Страница 184: ...174 Index Index...