92
Chapter 3
Technical Specifications
Noise Figure and Gain
Noise Figure and Gain
Performance is dependent on the ENR
1
of the noise source used:
1. Excess Noise Ratio
N8973A
N8974A (< 3.0 GHz), N8975A (< 3.0 GHz)
Noise Source ENR
4 - 7 dB
12 - 17 dB
20 - 22 dB
Noise Figure
Measurement Range
Instrument Uncertainty
0 to 20 dB
± < 0.05 dB
0 to 30 dB
± < 0.05 dB
0 to 35 dB
± < 0.1 dB
Gain
Measurement Range
-20 to +40 dB
Instrument Uncertainty
a
± < 0.17 dB
a. For measurement bandwidths below 4 MHz, and spacing between measurement points below
3 MHz, gain uncertainty may increase to a maximum of ±0.7 dB.
N8974A, N8975A (> 3.0 - 26.5 GHz)
Noise Source ENR
4 - 7 dB
12 - 17 dB
20 - 22 dB
Noise Figure
Measurement Range
Instrument Uncertainty
0 to 20 dB
± < 0.15 dB
0 to 30 dB
± < 0.15 dB
0 to 35 dB
± < 0.2 dB
Gain
Measurement Range
-20 to +40 dB
Instrument Uncertainty
a
± < 0.17 dB
a. For measurement bandwidths below 4 MHz, and spacing between measurement points below
3 MHz, gain uncertainty may increase to a maximum of ±0.7 dB.
Содержание NFA Series
Страница 25: ...13 2 Test Descriptions This chapter descibes each test required for calibration and performance verification...
Страница 114: ...104 Chapter3 Technical Specifications General Specifications...
Страница 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Страница 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Страница 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Страница 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Страница 184: ...174 Index Index...