NanoSuite Explorer D
Agilent Nano Indenter G200 User’s Guide
D-55
10
Click and drag a zoom box around the data at the bottom of the graph
as many times as you need to reveal the data.
11
Select
Save Sample As…
from the File menu.
12
Append the existing filename with
-mod
.
13
Click
Save
.
Exercise 4
Previously, you modified only how data are analyzed. In this exercise,
you will modify the test flow to change how the experiment is
performed. Beginning with the standard CSM method, you will create a
new method that uses a constant loading rate to get to a target force
quickly and then uses a constant strain rate to finish the experiment. You
will use 50 mN for the target force and 10 seconds for the ramp time.
You will also change the default oscillation amplitude to 10 nm.
1
Launch NanoSuite, logging in as a user with Creator priveleges.
2
Choose
Method > Open Method
.
3
Select
XP > G-Series CSM Standard Hardness, Modulus, and
Tip Cal
from the Filename drop-down menu.
4
Click
OK
.
5
Select
Save Method As…
from the Method menu.
6
In the Save Method File As dialog box, type in
G-Series CSM
Quick Ramp Then Constant Strain Rate
for the file name.
7
Select
Modified Methods
from the Save in drop-down menu.
8
Click
Save
.
9
Click the
Test
tab to reveal the Test page.
10
Click the
Define
tab to reveal the Define page.
11
Click the
Configuration
tab.
12
Click the
Inputs
folder.
13
In the Settings pane, right-click and select
Add Input…
from the
resulting pop-up menu.
14
In the Insert New Variable dialog box, enter the internal name
RampForceLimit
(no spaces).
15
Click
OK
.
Содержание Nano Indenter G200
Страница 1: ...Agilent Technologies Agilent Technologies Nano Indenter G200 User s Guide...
Страница 14: ...Agilent Nano Indenter G200 User s Guide 9 Express Test Tip Calibration H 12 Express Test Varied Force H 12...
Страница 96: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 4 Export Menu Mode Menu System Menu Help Menu...
Страница 109: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 17 Figure 4 18Load Review page configuration dialog box...
Страница 117: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 25 Figure 4 27Real Time Graph Setup dialog box...
Страница 121: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 29 Figure 4 31User Configuration dialog box...
Страница 136: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 44 Exit Use this icon to exit the software...
Страница 263: ...Commonly Used Procedures 6 Agilent Nano Indenter G200 User s Guide 6 44 Figure 6 62Flowchart 1 Figure 6 63Flowchart 2...
Страница 478: ...Agilent Technologies Agilent Nano Indenter G200 User s Guide Part Number G2A 13192 1 Rev C Agilent Technologies Inc 2012...