Commonly Used Procedures 6
Agilent Nano Indenter G200 User’s Guide
6-41
Figure 6-58
Confirm head, method and lead term settings in Area
Function Options tab
Once the calibration is complete, view the calibration outputs by
selecting the
Tip Area Results
or
was
chosen, the percent difference between the experimental area,
A(E)
, and
the calculated area from fit,
A
fit
, should be centered about zero and be
less than
5 %
for h
c
> 100 nm as shown in